Feasibility study for dose calculation with a radiation treatment planning system using a fixed-size electron cone
Su Chul Han1, Min Cheol Han2, Jihun Kim2
1Department of Radiation Oncology, Kangbuk Samsung Hospital, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
This study validates using commercial treatment planning systems (TPS) for small electron cone applicators. The method accurately calculates radiation doses, improving specialized electron therapy planning.
Area of Science:
- Medical Physics
- Radiation Oncology
- Dosimetry
Background:
- Commercial treatment planning systems (TPS) often lack native support for small, fixed-size electron cone applicators.
- Accurate radiation dose calculation is critical for effective and safe electron beam therapy.
Purpose of the Study:
- To assess the feasibility of using a commercial TPS for treatment planning with small fixed-size electron cone applicators.
- To validate the accuracy of monitor unit (MU) calculations and dose distributions generated by the TPS for these applicators.
Main Methods:
- Compared dosimetric characteristics (beam profiles, output factors) of a 6 MeV electron beam using a fixed-size cone versus a general applicator with cerrobend cutouts.
- Measured output factors and compared monitor units (MUs) from TPS with direct measurements.
- Validated TPS dose distributions against Gafchromic film measurements.
Main Results:
- Gamma analysis demonstrated high agreement for depth dose (>95% with 1%/1 mm) and beam profiles (>99% with 2%/2 mm).
- Indirect TPS calculations for the fixed-size electron cone applicator showed MU differences within 1.0% of direct measurements.
- 2D dose distribution agreement exceeded 99% (3%/3 mm) between film measurements and TPS calculations.
Conclusions:
- Commercial TPS can be feasibly adapted for treatment planning with small fixed-size electron cone applicators.
- The proposed method, incorporating relative output factors, provides a reliable approach for dose calculation in specialized electron therapy.