Study on the 241Am-Be neutron reference radiation field established at CAEP
Song Zhang1, Da-Jie Ding1, Ben-Jiang Mao1
1Institute of Nuclear Physics and Chemistry, China Academy of Engineering Physics, No. 64, Mianshan Road, Youxian District, Mianyang 621999, Sichuan, China.
The China Academy of Engineering Physics (CAEP) established a reliable americium-241/beryllium (241Am-Be) neutron radiation field for calibrating dose rate meters. This field meets ISO standards and provides accurate neutron energy spectrum and dose rate measurements.
Area of Science:
- Nuclear Physics
- Radiation Metrology
- Instrument Calibration
Background:
- The China Academy of Engineering Physics (CAEP) required a characterized neutron radiation field for instrument calibration.
- Existing facilities needed validation against established standards like ISO 8529.
Purpose of the Study:
- To evaluate the field characteristics of the 241Am-Be neutron reference radiation field at CAEP.
- To assess the instrument calibration capability within this established field.
- To ensure the field meets requirements for routine calibration and experiments.
Main Methods:
- Neutron energy spectrum analysis using multi-sphere spectrometer and Geant4 simulations.
- Quantity transmission verification via a precision long counter system between China Institute of Atomic Energy (CIAE) and CAEP.
- Calibration of three neutron dose rate meters from different manufacturers in both CAEP and CIAE fields.
- Uncertainty analysis of calibration factors for the calibrated instruments.
Main Results:
- The neutron radiation field at CAEP demonstrated acceptable room scatter, meeting ISO 8529 requirements.
- The measured neutron energy spectrum was deemed acceptable.
- The range and uncertainty of neutron ambient dose equivalent rate at CAEP were determined.
- Calibration factors and their uncertainties were found to be reasonable and reliable when validated across both fields.
Conclusions:
- The established 241Am-Be neutron radiation field at CAEP is suitable for routine calibration and experimental use.
- The field's characteristics and calibration capabilities have been successfully evaluated and validated.
- The study confirms the reliability of instruments calibrated within the CAEP neutron reference field.
More Related Videos
09:18Laser-heating and Radiance Spectrometry for the Study of Nuclear Materials in Conditions Simulating a Nuclear Power Plant Accident
Published on: December 14, 2017
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Related Concept Videos
Atomic Emission Spectroscopy: Overview
Atomic Emission Spectroscopy: Lab
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
Atomic Emission Spectroscopy: Instrumentation
Atomic Absorption Spectroscopy: Overview
When irradiated by EMR of a particular wavelength, these...