Performance of a New Tabletop Non-mydriatic Fundus Camera for Single-Field Diabetic Retinopathy Detection: A Pilot
Winston Padua1, Shalini Butola1, Suneetha Nithyanandam1
1Opthalmology, St. John's Medical College Hospital, Bengaluru, IND.
Cureus
|August 19, 2025
Summary
A new non-mydriatic tabletop fundus camera shows comparable performance to standard devices for diabetic retinopathy (DR) detection. This portable camera offers potential for point-of-care screening in diabetes mellitus patients.
Area of Science:
- Ophthalmology
- Medical Imaging
- Diabetology
Background:
- Diabetic retinopathy (DR) detection typically utilizes fundus cameras.
- Evidence for non-mydriatic fundus cameras with limited technical challenges in DR detection is scarce.
- This pilot study evaluates a novel non-mydriatic tabletop fundus camera for DR detection.
Purpose of the Study:
- To evaluate the performance of the Oivi fundus camera for DR detection.
- To compare its diagnostic accuracy against a standard reference device (Topcon NW400).
- To assess the feasibility of a single-field, macula-centered imaging approach.
Main Methods:
- 243 subjects with diabetes mellitus (DM) were recruited.
- Non-mydriatic, macula-centered images were captured using both Oivi and Topcon cameras.
- Two ophthalmologists independently graded images; discrepancies were resolved by consensus, with senior ophthalmologist's grading of standard camera images as ground truth.
Main Results:
- The Oivi camera demonstrated high diagnostic accuracy for DR, with sensitivity of 92.98% and specificity of 99.47%.
- DR detection rates were similar between the Oivi (23.86%) and standard (23%) cameras.
- The Oivi camera yielded a higher percentage of usable images (95.2%) compared to the standard camera (92.3%).
- Excellent inter-modality agreement (κ = 0.927) was observed for DR detection.
Conclusions:
- The Oivi tabletop fundus camera shows comparable performance to standard non-mydriatic devices for DR detection.
- Its portability and usability suggest potential for point-of-care screening in diabetic patients.
- Further large-scale studies are recommended to validate these findings for screening programs.