Related Experiment Video
Updated: May 9, 2026

Detection of Live Escherichia coli O157:H7 Cells by PMA-qPCR
Published on: February 1, 2014
Paper-based and chemical-free dual low-cost electrochemical biosensor for rapid detection of Escherichia coli O157:H7
Manman Du1,2,3, Junting Zhou2,4, Tianteng Hao2,4
1School of Environmental Science and Engineering, Tianjin University, Tianjin, 300072, China.
Abstract:
A low-cost electrical assay is reported using a paper-based electrochemical biosensor modified with carboxylated graphene oxide (GO-COOH), which is capable of directly immobilizing antibodies without chemical cross-linkers and is promising for the rapid detection of Escherichia coli (E. coli) O157:H7 in low-income areas. Antibodies were immobilized on the surface of the GO-COOH/C electrode at appropriate concentrations to directly and specifically capture targets. The GO-COOH/C paper-based electrochemical biosensor was capable of detecting as low as 0.26 CFU/mL of E. coli O157:H7 within 16 s selectively. The practicality of the GO-COOH/C paper-based electrochemical biosensor was evaluated using tap and river water samples. The detection results were similar to those of the standard plate culture method, but the analysis speed is much faster and has good cost-effectiveness (about US$ 1.81). The GO-COOH/C paper-based electrochemical biosensor stands out among various analytical techniques due to its dual low-cost advantages and excellent detection capabilities for E. coli O157:H7, which provide a strong guarantee for water quality safety in low-income areas.