A Validity and Reliability Study of the K-Muscle Controller Hand-Held Dynamometer in Measuring Shoulder's Internal
Stylianos Grigoriadis1, Vasileios Mylonas1, Thomas Nikodelis1
1Biomechanics Laboratory, School of Physical Education and Sport Science at Thessaloniki, Aristotle University of Thessaloniki, Thessaloniki, Greece.
Perceptual and Motor Skills
|August 21, 2025
Summary
The K-Muscle Controller is a reliable handheld dynamometer for measuring shoulder rotation strength in patients with shoulder impingement syndrome (SIS). It demonstrates high validity and reliability compared to existing methods.
Area of Science:
- Orthopedics
- Rehabilitation Medicine
- Biomedical Engineering
Background:
- Shoulder impingement syndrome (SIS) often requires accurate strength measurement for diagnosis and treatment.
- Existing dynamometers may have limitations in portability or ease of use.
- Validating new handheld devices is crucial for clinical practice.
Purpose of the Study:
- To assess the validity and reliability of the K-Muscle Controller, a new handheld dynamometer.
- To compare its measurements of shoulder external and internal rotation strength against the MicroFET2 device.
- To evaluate its utility in patients diagnosed with shoulder impingement syndrome (SIS).
Main Methods:
- Two groups of SIS patients (n=22 and n=25) participated.
- Isometric shoulder rotation strength was measured using the K-Muscle Controller and/or MicroFET2.
- Intra-Class Correlation (ICC), Coefficient of Variation (CoV), and Bland-Altman plots were used for analysis.
Main Results:
- High ICC values (0.944–0.986) indicated excellent reliability and validity for the K-Muscle Controller.
- CoV values ranged from 31.77% to 48.57%.
- Bland-Altman plots confirmed strong agreement between measurements without systematic bias.
Conclusions:
- The K-Muscle Controller is a reliable tool for measuring shoulder external and internal rotation strength.
- Its performance is comparable to established devices like the MicroFET2.
- This device shows promise for clinical use in assessing shoulder strength in both healthy and pathological conditions.
Keywords:
hand-held dynamometryreliabilityshoulder external rotationshoulder internal rotationvalidity