Evaluation of measurement uncertainty results calculated by two different methods and total analytical error for
Mehmet Akif Bozdayi1, İsmet Gamze Bozdayı2, Gökhan Çakırca1
1Mehmet Akif Inan Education and Research Hospital, Department of Medical Biochemistry, Sanliurfa, Türkiye.
Background:
The analytical performance metrics of ethanol testing are critically important due to their legal implications when presenting and interpreting results. Measurement uncertainty (MU) and total analytical error (TAE) are essential approaches for evaluating and improving the quality of measurement procedures. This study aimed to calculate MU and TAE values, which assess the reliability of ethanol test results from different perspectives, and to evaluate the impact of MU values, calculated using two different methods, on the legal threshold.
Methods:
MU values were calculated following the guidelines of Nordtest and ISO/TS 20914. TAE was determined using the formula TAE%=1.65×CV%+Bias%. External and internal quality data from ethanol testing conducted between July 1, 2022, and June 30, 2024, were used for calculations.
Results:
The expanded MU values for ethanol testing were 13.95% according to the Nordtest Guide, 10.94% for low level and 9.59% for high level according to the ISO/TS 20914 Guide. The calculated TAE values were 12.59 for low levels, 11.47 for high levels, and 12.57 overall. MU and TAE values for ethanol testing in our laboratory remained within the allowable total error (±20%) according to CLIA 2024.
Conclusions:
We believe each laboratory should report ethanol test results and their respective MU values, particularly when evaluating results close to legal thresholds. Furthermore, we suggest that scientific committees standardise the method for calculating MU and define a target limit.
Related Concept Videos
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Uncertainty: Overview
Uncertainty in Measurement: Accuracy and Precision
Data Validation
Key parameters for method validation include:
Uncertainty in Measurement: Reading Instruments
Contaminants and Errors
Another key consideration is determining the appropriate number of samples required to...