Development and evaluation of a comprehensive QA phantom for image quality and geometrical accuracy in CT and CBCT
Minsoo Chun1,2, Jaewon Hong1, Jin Hwa Choi3
1Department of Radiation Oncology, Chung-Ang University Gwang Myeong Hospital, Gyeonggi-do, Republic of Korea.
Purpose:
This study presents MZ CUBE, a newly developed multipurpose phantom designed to assess both image quality and geometrical accuracy in image-guided radiotherapy.
Methods:
The phantom is a 16 × 16 × 16 cm3 cube made of acrylonitrile butadiene styrene and designed to evaluate HU constancy, HU uniformity, spatial integrity, spatial resolution, low contrast resolution, and geometrical accuracy. Two CT simulators (Big Bore RT and Discovery RT Gen3) and two cone-beam CT (CBCT) systems (XVI and OBI) were used for verification alongside commercial phantoms (ACR 464 and CatPhan 504).
Results:
HU constancy differences were within 3 HU, and uniformity was within 1 HU for CT and 2 HU for CBCT, based on repeated MZ CUBE measurements. Spatial resolution with the MZ CUBE was 7 lp/cm (Big Bore RT), 8 lp/cm (Discovery RT Gen3), 7 lp/cm (XVI), and 8 lp/cm (OBI). The minimum discernible disk diameters were 7, 5, 6, and 6 mm, respectively. Geometrical accuracy was confirmed with translational offsets below 1 mm and rotational deviations below 0.2° for both CBCT systems.
Conclusions:
MZ CUBE enables comprehensive image quality and geometric accuracy evaluation in a single scan, offering an efficient and cost-effective solution for radiotherapy QA.
More Related Videos
09:57Development and Evaluation of 3D-Printed Cardiovascular Phantoms for Interventional Planning and Training
Published on: January 18, 2021
08:41Patient-Specific Polyvinyl Alcohol Phantom Fabrication with Ultrasound and X-Ray Contrast for Brain Tumor Surgery Planning
Published on: July 14, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography