A model-based method for reporting mammographic diagnostic reference levels for any compressed breast thickness: A
Ahmad A Alhulail1, Salman M Albeshan2, Maha M Almuqbil3
1Department of Radiology and Medical Imaging, College of Applied Medical Sciences, Prince Sattam bin Abdulaziz University, Al-Kharj, Saudi Arabia.
A new equation-based approach provides precise diagnostic reference levels (DRLs) for mammography, improving radiation dose optimization beyond conventional methods. This method offers continuous DRL values for any compressed breast thickness (CBT), enhancing patient safety in breast cancer screening.
Area of Science:
- Medical Imaging
- Radiological Physics
- Radiation Safety
Background:
- Mammography is vital for early breast cancer detection, but involves ionizing radiation requiring dose monitoring.
- Conventional diagnostic reference levels (DRLs) use broad compressed breast thickness (CBT) ranges, limiting precise radiation dose optimization.
- Individual variations in CBT hinder the effectiveness of current DRL reporting methods.
Purpose of the Study:
- To develop a novel equation-based method for determining diagnostic reference levels (DRLs) applicable to any compressed breast thickness (CBT).
- To enhance the precision of mammographic radiation dose optimization by addressing individual patient variations.
Main Methods:
- Utilized the 75th percentile of median average glandular dose (AGD) from 187,704 mammograms across nine centers.
- Compared three DRL estimation approaches: a simple range, DRLs per 10-mm CBT ranges, and a proposed DRL equation derived from curve fitting.
- Assessed the reliability and differences in DRL estimations across the evaluated methods.
Main Results:
- A bi-exponential equation with high reliability (R² > 0.98) was established for DRL estimation based on CBT.
- The equation approach yields continuous DRL values for any specific CBT.
- Significant discrepancies were observed, with the equation approach showing potential dose differences exceeding 35% and 175% compared to range-based methods.
Conclusions:
- The proposed DRL equation approach offers superior precision for any CBT compared to conventional range-based methods.
- This equation-based DRL system can automate mammographic dose optimization when integrated by vendors.
- The findings support a more individualized and effective approach to radiation safety in mammography.