Evaluation of size-specific dose estimates in CT kidneys, ureters, and bladder scans: comparison of central-slice and
Khaled Alenazi1, Abdullah Abuhaimed2, Ali Alanazi3
1Radiological Sciences Department, College of Applied Medical Sciences, King Saud University, Riyadh 4545, Saudi Arabia.
Objectives:
To evaluate the size-specific dose estimates (SSDE) in adult patients undergoing kidneys, ureters, and bladder (CT KUB) scans using 2 approaches: water-equivalent diameter (Dw) and effective diameter (Deff). The study also aimed to assess the accuracy of using a single central image slice for SSDE estimation.
Methodology:
Ethical approval was obtained to collect patient data from a local hospital. CT images from 203 adult patients were retrieved and processed using IndoseCT software to calculate patient size metrics and corresponding SSDE values.
Results:
SSDE values calculated using the Dw,mean and Deff,mean were comparable: 10.3 ± 2.9 mGy and 10.2 ± 2.9 mGy, respectively. When using a single central slice, the mean percentage differences were -1.8 ± 3.9% for Dw and -1.1 ± 4.6% for Deff, with all values falling within ±11%. A strong correlation was observed between SSDE values derived from single-slice and full-slice measurements (R2 > 0.97), showing slightly better agreement for Dw.
Conclusion:
While Dw offers a more accurate estimation of patient size, the minimal differences observed suggest that Deff is a suitable alternative when Dw specific tools are unavailable. Additionally, using a single central slice is a practical and efficient method to estimate SSDE, significantly reducing computational demands.
Advances In Knowledge:
This study provides clinical validation that SSDE can be reliably estimated using a single-slice method in CT KUB examinations, offering a substantial reduction in processing time. It also demonstrates that Deff is a viable substitute for Dw when access to advanced imaging analysis tools is limited.
More Related Videos
Related Concept Videos
Imaging Studies I: Kidney, Ureter, and Bladder Studies
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...