A methodological approach for the comparison of CT systems in terms of radiation dose and image quality
Evgenia Konstantinou1, Antonios Ε Papadakis2, Maria Daskalogiannaki3
1Department of Medical Physics, Medical School, Department, University of Crete 71003 Heraklion, Crete, Greece.
Purpose:
This study aims to apply a comparative methodology for two different computed tomography (CT) scanners, by evaluating patient radiation dose and image quality.
Materials & Methods:
A total of 189 consecutive non-enhanced and enhanced abdominal examinations, were performed using General Electric Revolution GSI (scanner A) and Siemens Somatom Drive (scanner B) scanners. Both protocols had been previously optimized by the same team for the two scanners, ensuring consistent image quality during comparison. CT dose index volume (CTDIvol) and dose-length product (DLP) were recorded, and percent dose difference between scanners was estimated. Image quality was objectively assessed using image noise and contrast-to-noise ratio (CNR), and subjectively with a five-point scale. Utilizing the same protocols, an anthropomorphic phantom was irradiated to estimate organ doses. Statistical analysis was conducted to compare the examined parameters.
Results:
The CTDIvol and DLP for two scanners were 16.4-21.8 mGy and 494.7-1030.3 mGy*cm, respectively. Doses for scanner B were up to 41 % lower than scanner A. No significant CTDIvol and DLP differences were found for unenhanced scans. Organ doses ranged from 5.0 to 16.9 mGy for both scanners, with scanner B delivering lower doses. Image quality was comparable between two CT systems. No statistical differences were found for image quality parameters, except for CNR in non-enhanced examinations. Radiologists' ratings were consistent with the objective assessment.
Conclusion:
A methodology was applied to compare two different CT scanners, regardless of patient selection criteria. Scanner B achieved lower doses for contrast-enhanced exams than scanner A.
More Related Videos
08:30X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
08:02Author Spotlight: Enhanced Quantification of Cardiovascular Calcification Progression for Longitudinal Micro PET/CT Studies in Small Research Animals
Published on: November 15, 2024
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography