Characterization and Clinical Translation of a Novel Prototype Kilovoltage Dual-Layer Imager for Onboard Imaging
Thomas C Harris1, Matthew Jacobson1, Raphael Bruegger2
1Department of Radiation Oncology, Brigham and Women's Hospital, Dana Farber Cancer Institute and Harvard Medical School, Boston, Massachusetts.
Purpose:
Dual-energy imaging offers several benefits, including artifact reduction and material decomposition. This imaging may be achieved via multiple detecting layers with beam hardening between layers creating spectral separation. We describe the construction, characterization, and clinical translation of a novel prototype dual-layer flat-panel detector for kilovoltage (kV) onboard radiation therapy imaging.
Methods And Materials:
The dual-layer imager (DLI) was designed with the first layer matching existing detector construction: cesium-iodide (CsI) scintillator and amorphous silicon thin-film transistor photodiode array. The second layer has a slightly thicker CsI scintillator, providing extra photon detection efficiency. The prototype DLI was built by an industry partner. For clinical deployment, top layer information was sent to the treatment console; data from both layers were sent to a research PC for retrospective analysis. Modulation transfer function (MTF(f)) and noise power spectrum were measured for the top, bottom, and combined layers. Detective quantum efficiency (DQE(f)) was calculated from the results. Leeds phantom imaging further assessed detector performance. Initial patient data were analyzed to determine material decomposition feasibility. Log-weighted subtraction and virtual monoenergetic images were generated.
Results:
The DLI was deployed on a clinical TrueBeam linear accelerator (linac). For 120 kVp, the DQE(0) is 0.48 for the top layer, 0.13 for the bottom layer (using initial fluence), and 0.61 for the combined. Leeds contrast-to-noise ratio, with the top layer as the baseline, was -45.8% for the bottom layer and +12.4% for the combined. MTF50 went from 1.38 top layer to 0.96 combined. Initial log-weighted subtraction of layers of patients' lung kV's successfully suppressed bone material. Virtual monoenergetic images were generated: 40 keV increased soft tissue contrast by 1.6×, whereas 120 keV increased uniformity in metal artifact regions by 41.4%.
Conclusions:
A novel kV DLI was constructed and translated on a clinical linac. Combining the layers increases DQE with an MTF reduction. More importantly, it adds dual-energy imaging capabilities to a linac. Preliminary results show promise for spectral imaging applications, such as metal artifact reduction and bone removal to enable better lung tumor visualization.
More Related Videos
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
X-ray Imaging
Imaging Studies III: Computed Tomography
Imaging Studies II: Ultrasonography
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...