A Precision Medicine Tool for Measurement and Monitoring of Hemoglobin S in Sickle Cell Disease Patients Receiving Transfusion Therapy
1.8K
Hb Tacoma is difficult to detect and may cause false results with different HbA1c methods
Britta Landin1,2, Soheir Beshara1,3, Johan Eskils1
1Department of Clinical Chemistry, Karolinska University Hospital, Stockholm, Sweden.
Scandinavian Journal of Clinical and Laboratory Investigation
|August 25, 2025
Summary
The haemoglobin variant Hb Tacoma can be misinterpreted by certain HbA1c testing methods, leading to inaccurate results. Immunoassays and some chromatography methods accurately measure HbA1c in individuals with Hb Tacoma.
Area of Science:
- Clinical Chemistry
- Haematology
- Biochemistry
Background:
- The haemoglobin variant Hb Tacoma (HBB:c.93G>T;CD30(Arg>Ser)) is found in about 1 in 1000 patients undergoing HbA1c testing.
- Homozygous individuals for Hb Tacoma show no severe hematological issues.
Purpose of the Study:
- To evaluate the impact of Hb Tacoma on various HbA1c measurement methods.
- To identify reliable methods for accurate HbA1c quantification in the presence of Hb Tacoma.
Main Methods:
- Comparison of HbA1c results from different analytical platforms including cation exchange chromatography, immunoassays, affinity, and enzymatic methods.
- Investigation of chromatogram interpretation for Hb Tacoma heterozygotes using the Bio-Rad Variant II system.
Main Results:
- The Bio-Rad Variant II method showed significantly lower HbA1c values (approx. 40% lower) due to coelution of glycated Hb Tacoma.
- Immunoassays (Roche Tina-quant Gen. 3, Siemens DCA Vantage), affinity, and enzymatic methods provided accurate HbA1c results.
- Bio-Rad D-100 and Tosoh G11 methods showed unpredictable deviations or positive bias.
- Sebia Capillarys 3 Tera capillary electrophoresis exhibited high imprecision with Hb Tacoma samples.
Conclusions:
- Certain HbA1c testing methods, particularly cation exchange chromatography like Bio-Rad Variant II, can lead to misinterpretation and inaccurate results in the presence of Hb Tacoma.
- Immunoassays and other specific methods demonstrate reliability for HbA1c measurement in individuals with this haemoglobin variant.
Keywords:
Glycated hemoglobinchromatography, ion exchangeelectrophoresis, capillaryhemoglobin Tacomahemoglobinopathiesimmunoassaypoint-of-care testing