Validation of a portable three-dimensional imaging system for volumetric measurement in the periorbital region
Xuan Zhang1, Ji Shao1, Ningxin Dai1
1Eye Center of the Second Affiliated Hospital, School of Medicine, Zhejiang University, Zhejiang Provincial Key Laboratory of Ophthalmology, Zhejiang Provincial Clinical Research Center for Eye Diseases, Zhejiang Provincial Engineering Institute on Eye Diseases, Hangzhou, China.
Objectives:
To evaluate the reliability and accuracy of a portable structured light-based three-dimensional (3D) imaging system for volumetric measurement in the periorbital region.
Methods:
Five hemispherical 3D printed resin models with different sizes and colors, including the flesh-colored models with 6 (Model 1), 4 (Model 2), and 2 (Model 3) mm in diameter, and the black (Model 4) and gray (Model 5) models with 6 mm in diameter, were affixed to the lower eyelid or upper eyelid of 40 volunteers. One researcher used the portable 3D imaging system to collect the 3D images and two raters measured the volume of the models on images. Intra-device, intra-rater and inter-rater reliabilities and accuracy of the volumetric measurement were evaluated by intraclass correlation coefficient (ICC), mean absolute difference (MAD), technical error of measurement (TEM), relative error measurement (REM), and relative TEM (rTEM).
Results:
The intra-device reliability of the 3D imaging system for volumetric measurement in the periorbital region was excellent (ICC = 0.922, MAD = 0.11 mm3, TEM = 0.09 mm3, REM = 0.19%, rTEM = 0.15%). The intra-rater reliability for volumetric measurement of the Model 1 on the lower eyelid was the highest (ICC = 0.953, MAD = 0.08 mm3, TEM = 0.06 mm3, REM = 0.13%, rTEM = 0.11%). The inter-rater reliability for volumetric measurement of the Model 1 on the lower eyelid was the highest (ICC = 0.889, MAD = 0.14 mm3, TEM = 0.11 mm3, REM = 0.24%, rTEM = 0.19%). The accuracy for volumetric measurement of the Model 1 on the lower eyelid was the highest (MAD = 0.24 mm3, REM = 0.43%).
Conclusion:
The portable 3D imaging system proved to measure the volumes of periorbital flesh-colored objects reliably and accurately. This finding demonstrated that this device has great potential for diagnosis, post-operative evaluation, and long-term follow-up of volumetric changes in oculoplastics.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Three-Dimensional Force System
Three-Dimensional Force System:Problem Solving
To solve a three-dimensional force system, first resolve each force into its respective scalar components. Do this using...
Estimation of the Physical Quantities
Imaging Studies III: Computed Tomography
Finding Volume Using Cross-Sectional Area