Gradually increasing pacing impedances with normal sensing values in pacemaker and defibrillation leads: A watchful
Eva Roseboom1, Fenna Daniëls2, Martijn G Klingenberg1
1Department of Cardiology, University of Groningen, University Medical Center Groningen, Groningen, The Netherlands.
Background:
The increasing prevalence of cardiac implantable electronic devices (CIEDs) has led to greater data collection, including pacing impedance, a key measure of lead integrity and electrical function. Although acute impedance changes may signal lead fractures or insulation defects, gradual increases are often linked to tissue-interface calcification. The optimal management of patients with progressively rising impedance remains uncertain.
Objective:
This study aimed to assess the prevalence, clinical consequences, and management strategies for CIED leads showing a gradual increase in pacing impedance of >1200 Ω.
Methods:
We conducted a single-center retrospective cohort study of 59 patients (42 with an implantable cardioverter-defibrillator and 17 with a pacemaker) over a median follow-up of 9.2 years (range 2.8-18.5). Data on patient demographics, device characteristics, and outcomes associated with elevated impedance were analyzed. Impedance values, intracardiac signals, and pacing thresholds were assessed at baseline and every 6 months up to 18 years.
Results:
Among patients with an implantable cardioverter-defibrillator, 33 of 42 leads (78.6%) were preserved, whereas 9 were replaced (5 prophylactically, 2 owing to oversensing, and 2 for threshold issues). In patients with a pacemaker, 12 of 17 leads (70.6%) were preserved, with 5 replacements (4 prophylactic and 1 for threshold issues).
Conclusion:
A watchful waiting strategy seems safe for patients without sensing or threshold concerns and in the absence of a pacing indication. A patient-centered approach should guide the management of gradual impedance increases in CIED leads.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Dysrhythmias VI: Management of Dysrhythmias
Electrophysiology of Normal Cardiac Rhythm
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Decreased pulse rate
There are specific risk factors that can elevate the likelihood of developing bradycardia. Advanced age is a significant factor, with...
Dysrhythmias III: Characteristics of Dysrhythmias