Related Experiment Video
Updated: Sep 10, 2025

Laryngeal Mask Airway LMA Placement in a Neonatal Patient Simulator Using a Non-Inflatable Supraglottic Airway SGA
Published on: July 14, 2023
Applied forces with partially vs. fully inflated face masks during neonatal ventilation: a randomized crossover
Francesco Cavallin1, Andrea Zamunaro2, Sabina Maglio3,4
1Independent statistician, Solagna, Italy.
Background:
Positive pressure ventilation with partially (25 ml) vs. fully (35 ml) inflated face mask was compared in terms of applied forces on a newborn manikin face, cuff pressure and air leak.
Methods:
A crossover randomized controlled trial of face mask ventilation with partially (25 ml) vs. fully (35 ml) inflated face mask in a term neonatal manikin. The trial involved 28 neonatologists and pediatric residents. The primary outcome was the force applied to the manikin face, while the secondary outcomes were cuff pressure, ventilation time with leak <25% around the mask, and participants' view on struggle in delivering effective ventilation, fatigue in achieving good mask seal, and satisfaction about the mask.
Results:
Chin and cheekbone sensors measured higher top 10th percentile of applied forces with partially vs. fully inflated face mask. Median and top 10th percentile of cuff pressures were higher with the partially inflated mask. Ventilation time with mask leak <25% was not statistically different using partially vs. fully inflated face masks, and participants reported similar opinions about the two procedures.
Conclusions:
The lower force applied during positive pressure ventilation with a fully inflated face mask in a neonatal manikin model may be an appealing target, nonetheless the real-life benefits should be addressed in future investigations.
Registration:
clinicaltrial.gov NCT0616138.
Impact:
Health care providers may apply different forces to improve mask seal during neonatal resuscitation. Still, data regarding applied forces with various face-mask air cushion volumes is unknown. In a neonatal manikin model, using a partially inflated (25 ml) face mask during PPV was associated to higher applied forces and cuff pressures compared to a fully inflated (35 ml) face mask. The participants reported comparable opinions about the two air cushion volumes. Applying less force on neonatal face might be a desirable objective, nonetheless future investigations should assess its real-life impact.
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...