Ability of Turbine-Based Ventilators to Maintain Tidal Volume Following Transient Conditions
Vincent Delord1, Isabelle Vaugier2, Hélène Prigent1,2,3
1Mr. Delord, Prof. Prigent, and Prof. Lofaso are affiliated with Department of Clinical Physiology, AP-HP, Hôpital Raymond-Poincaré, Université de Versailles Saint Quentin en Yvelines, Garches, France.
Background:
This study aimed to evaluate the efficacy and reliability of maintaining a constant tidal volume (VT) using a square-wave flow pattern in currently available turbine-based ventilators (TBVs) compared with a bellows-based ventilator (BBV), under conditions of abrupt changes in ventilator load. Such changes frequently occur in ventilator-dependent patients during mouthpiece ventilation and sleep with a mask, including events such as air stacking (AS), leakage, disconnection for speaking or eating/drinking, and complete obstruction.
Methods:
Five TBVs and one BBV were assessed in a bench test simulating 4 distinct transient events (AS, leakage, disconnection, and complete obstruction) across 3 consecutive cycles.
Results:
During AS, only one TBV maintained VT as effectively as the BBV, whereas the others showed a significant decrease in VT when the elastic load increased because of the preceding stacked breath(s). During leakage, the same TBV again maintained VT as effectively as the BBV, both during and after the leak. In contrast, the other TBVs exhibited a transient increase in delivered VT at the onset of leakage, followed by a decrease once it ended. Except for the BBV, disconnection resulted in continuous air flow, with flow peaks occurring at the moments when VT was expected to be delivered. Following the termination of obstructive apnea, all devices maintained the targeted VT, except for one BBV, which delivered a VT above baseline during the 2 subsequent cycles.
Conclusions:
At least one BBV consistently maintained a stable VT under various conditions, including AS, leakage, postapnea recovery, and postdisconnection. However, disconnection was associated with continuous air flow in the TBVs, which could be disruptive during mouthpiece ventilation disconnection for speaking, eating, or drinking. This unintended effect warrants further investigation.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Respiratory Volumes and Capacities
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...