Minimizing high sensitivity troponin T delta variation at low concentration using BD Barricor blood collection tube
Albert K Y Tsui1, Jialin Qiu2, Isolde Seiden-Long3
1Department of Laboratory Medicine and Pathology, College of Health Science, Faculty of Medicine and Dentistry, University of Alberta, Edmonton, Alberta, Canada; Alberta Precision Laboratories, Alberta, Canada.
Objective:
Reproducible low troponin concentrations from high-sensitivity troponin (hs-cTn) assays are paramount to accurate risk determination in the accelerated diagnostic pathway. Total variation consists of pre-analytical, analytical and biological components. While analytical and biological variations cannot be readily modifiable, minimizing pre-analytical variation is desirable and potentially attainable. The BD Barricor collection tube has previously been demonstrated to reduce pre-analytical variation in test results. The goal of the study is to determine whether BD Barricor tubes provide more reproducible hs-cTnT results compared to plasma separator tubes (PST) at concentrations ≤ 20 ng/L.
Methods:
Paired intra-patient hs-cTnT results collected less than 1 h apart in the emergency department were retrospectively analyzed from nine urban hospitals which primarily use either PST (n = 336 pairs) or Barricor (n = 327 pairs) collection tubes for troponin. Total variation of the replicated measurements was calculated for hs-cTnT ≤ 20 ng/L. The numbers of paired intra-patient samples were grouped based on decisive absolute delta thresholds as indicated by the European Society of Cardiology 0/1 h algorithm; delta < 3 ng/L, 3-4 ng/L, ≥5 ng/L.
Results:
The total testing variation for hs-cTnT collected in PST is 14.8 % while Barricor is 8.6 % for hs-cTnT ≤ 20 ng/L. The proportion of delta values < 3 ng/L between the intra-patient replicates is 80.4 % (95 % CI: 75.7-84.5 %) in PST compared to 95.4 % (95 % CI: 92.5-97.4 %) in Barricor (p < 0.001). Median time for serial sampling in PST is 41 min (IQR:18-53) and Barricor is 45 min (IQR 23-54).
Conclusion:
The use of Barricor tubes demonstrated reproducible and less variable hs-cTnT replicates at concentration ≤ 20 ng/L when compared to a hospital that does not use Barricor tubes.
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Therapeutic Drug Monitoring: Drug Analysis Methods
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...