Related Experiment Video
Updated: Sep 9, 2025

Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
Published on: March 22, 2022
Performance comparison of Mindray EH-2090 Urinalysis analyzers and Sysmex UF-5000
Juan Chen1, Chen Zhang2, Qiaoping Xu1
1Department of Laboratory Medicine, Peking University Shenzhen Hospital, Shenzhen, China.
Abstract:
In this study, we used manual microscopy as the gold standard, and compared the analytical and clinical performance of EH-2090 and UF-5000 to evaluate their application characteristics in daily clinical practice. Results show the EH-2090 has comparable analytical performance to Sysmex UF-5000. In clinical performance, the EH-2090 shows better counting accuracy for urine formed elements except bacteria. Besides, the EH-2090 could provide more accurate assessment especially in samples with yeast cells, crystals and dysmorphic red blood cells.