Pre-analytical interference in point-of-care troponin T testing: a case series
Lekha Priyadharshini Kamarajan1, Priyanshu Tripathi1, Sushil Kumar1
1Department of Biochemistry, All India Institute of Medical Science Patna, Patna, India.
Objectives:
To examine the impact of haematological and biochemical abnormalities on the failure of point-of-care troponin T (POCT) testing in patients with suspected acute coronary syndrome (ACS).
Case Presentation:
Five patients underwent Roche Troponin T Card Test (lateral flow immunoassay) in an emergency setting. Despite correct sampling and procedural adherence, no valid results were obtained resulting in an aborted test. Laboratory analysis revealed severe anaemia, polycythaemia, leucocytosis, thrombocytopenia, and hepatic and renal dysfunction across cases. After stabilization, repeat POCT yielded valid results in all survivors, correlating with normalized haematological parameters.
Conclusions:
Pre-analytical factors such as extreme haematocrit, leucocytosis, and biochemical derangements can cause POCT failure. Pre-testing screening and guideline updates are essential to optimize POCT reliability in acute care.
More Related Videos
10:12A Sensitive and Specific Quantitation Method for Determination of Serum Cardiac Myosin Binding Protein-C by Electrochemiluminescence Immunoassay
Published on: August 8, 2013
06:16Signal Acquisition, Score Interpretation, and Economics of a Non-Invasive Point-of-Care Test for Coronary Artery Disease
Published on: August 9, 2024
Related Concept Videos
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...
Acute Coronary Syndrome III: Diagnostic Studies
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Acute Coronary Syndrome I: Introduction
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
Myocarditis II: Clinical Features and Diagnostic Tests