Threaded K-wire vs cortical screw fixation in O'Driscoll type 2 and 3 coronoid fractures: a comparative biomechanical
Cafer Ozgur Hancerli1, Ada Ozcecelik2, Ergun Bozdag2
1Department of Orthopedics and Traumatology, Bahçeşehir University Faculty of Medicine, Istanbul, Turkey.
Background:
Coronoid fractures significantly impact elbow stability, yet limited biomechanical data exists comparing fixation methods for different fracture types. This study aimed to compare the biomechanical performance of threaded K-wire versus cortical screw fixation in O'Driscoll type 2 and 3 coronoid fractures.
Methods:
Twenty-eight synthetic ulnar bones were divided into four groups (n = 7 each): Type 2 with K-wire fixation, Type 2 with screw fixation, Type 3 with K-wire fixation, and Type 3 with screw fixation. Fractures were created, reduced, and fixed under fluoroscopic guidance. Specimens underwent biomechanical testing using a custom-made apparatus to evaluate load to failure (N), displacement (mm), and stiffness (N/mm). Two-way ANOVA and post-hoc Tukey's tests were used for statistical analysis.
Results:
Type 2 fractures with screw fixation demonstrated the highest load to failure (1392.59 ± 76.77 N), followed by Type 2 with K-wire fixation (1155.00 ± 200.81 N), Type 3 with K-wire fixation (1093.65 ± 248.68 N), and Type 3 with screw fixation (1058.54 ± 320.46 N), though differences were not statistically significant (p = 0.086). For stiffness, Type 2 fracture fixation fractures exhibited significantly higher values (∼256 N/mm) compared to Type 3 fractures (∼160 N/mm) regardless of fixation method (p = 0.002, Cohen's d = 1.55). The fixation method itself (K-wire vs. screw) did not significantly affect any biomechanical parameter (p > 0.05).
Conclusion:
O'Driscoll Type 2 fracture fixation provide superior biomechanical stability compared to Type 3 fractures, primarily through enhanced stiffness. While Type 2 screw fixation demonstrated the highest load to failure values, K-wire fixation in Type 2 fractures offered comparable stiffness. These findings suggest that fracture type has a more profound impact on mechanical performance than the choice between K-wire and screw fixation, giving surgeons flexibility in fixation choice for Type 2 fractures while maintaining adequate stability for early rehabilitation.
More Related Videos
04:41Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025
04:19Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024
Related Concept Videos
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the procedure...
Normal Strain under Axial Loading
Relation between Poisson's ratio, Modulus of Elasticity and Modulus of Rigidity