Evaluation of flow control using PID versus fuzzy logic in an electropneumatic circuit for pulmonary ventilation
Lina González1, Issa Griffith1, Alfredo Lescher1
1Centro I+D+i de Biotecnología, Energías Verdes y Cambio Climático (BEVCC), Laboratorio de Investigación Experimental de Bioseñales, Biomedical Engineering, Faculty of Biosciences and Public Health, Universidad Especializada de las Americas (UDELAS), Albrook, Paseo de La Iguana, Panama.
Abstract:
High-tech mechanical ventilators are engineered to deliver precise and consistent airflow, which is critical for effective respiratory therapy. This study evaluates flow control performance in a custom-built electro-pneumatic ventilator prototype, comparing Proportional-Integral-Derivative (PID) control with Fuzzy Logic Control (FLC) through real-time experiments on a test-lung platform to assess accuracy and adaptability under dynamic conditions. A laboratory based experimental study was conducted under laboratory conditions, using a test lung simulator and real-time flow data acquisition. The analysis included time-domain performance metrics and statistical validation through Bland-Altman analysis. Results indicate that both controllers meet the accuracy thresholds expected in commercial systems. However, the fuzzy logic controller exhibited narrower limits of agreement and lower standard deviation, indicating greater consistency. While PID control responded faster, with a settling time between 0.32 and 0.43 seconds, FLC achieved superior performance in high-demand scenarios, delivering an entire volume of 900 mL. Stability analysis using the Jury Test and Nyquist criteria confirmed that both systems are dynamically stable. Notably, the FLC curve in the Nyquist plot remained farther from the critical point (-1, 0j), indicating enhanced robustness against disturbances. These findings suggest that FLC may offer a reliable alternative to PID in nonlinear ventilation scenarios, particularly in resource-constrained environments seeking technological autonomy.
More Related Videos
08:35Interactive and Visualized Online Experimentation System for Engineering Education and Research
Published on: November 24, 2021
06:45Design and Application of a Fault Detection Method Based on Adaptive Filters and Rotational Speed Estimation for an Electro-Hydrostatic Actuator
Published on: October 28, 2022
Related Concept Videos
PID Controller
PI Controller: Design
Time-Domain Interpretation of PD Control
Consider the example of control of motor torque. Initially, a positive...
Time and frequency -Domain Interpretation of PI Control
Acting as a low-pass filter, the PI controller slows the system's response and extends settling times. This requires...
PD Controller: Design
Designing a continuous-data controller requires selecting and linking components like adders and integrators, which are fundamental in Proportional,...
Frequency-Domain Interpretation of PD Control
The proportional control gain, combined with the...