K-Edge Imaging Using a Clinical Dual-Source Photon-Counting CT System
Martin Vicente Rybertt1,2, Leening P Liu1, Pooyan Sahbaee3
1Department of Radiology, University of Pennsylvania, Philadelphia, PA 19104, USA.
Purpose:
To evaluate the feasibility and performance of K-edge imaging of iodine (I) and gadolinium (Gd) on a clinically available photon-counting computed tomography (PCCT) system.
Methods:
A dual-source clinical PCCT scanner with four energy thresholds (20, 55, 72, 90 keV) was used to scan phantoms containing pure and mixed solutions of I and Gd across multiple concentrations (1-10 mg/mL) and radiation doses (1-8 mGy). Multi-material decomposition was performed using a calibration-based, image-domain algorithm to generate material-specific maps. Quantitative accuracy was assessed using Bland-Altman analysis and contrast-to-noise ratio (CNR), while noise and bias trends were statistically analyzed using non-parametric tests.
Results:
K-edge imaging was successfully achieved on a clinical PCCT system with accurate decomposition of I and Gd across varying concentrations, solution types (pure/mixed), and dose levels. Quantitative bias was significantly influenced by radiation dose, concentration, and solution type (p < 0.0004). Increased radiation dose and contrast concentration improved quantification accuracy, with maximum bias reductions of 0.9 (I) and 0.3 mg/mL (Gd). CNR correlated linearly with concentration (R2 > 0.99) and moderately with dose (R2 = 0.85-0.94), achieving peak values of 13 (I) and 16 (Gd) at 8 mGy. Mixed solutions showed reduced performance compared to pure solutions, i.e., CNR of 5 mg/mL Gd solutions increased by 0.6 per mGy in pure solutions while by 0.5 per mGy in mixtures. Noise was dependent on dose but not on concentration or solution type.
Conclusion:
This study establishes the feasibility of K-edge imaging using a clinical PCCT system and demonstrates accurate, simultaneous decomposition of I and Gd in pure and mixed solutions. These findings support the clinical translation of K-edge imaging and highlight PCCT's potential for advanced dual-contrast and molecular imaging applications.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation I: X-ray and CT
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...