Validation status of digital blood pressure monitors in Peru: a comprehensive assessment with a validation-process
Fabian A Chavez-Ecos1,2, Sandra S Chavez-Malpartida3, Alexander M Parra-Huaroto1
1Facultad de Medicina Humana, Universidad Nacional San Luis Gonzaga, Ica.
Background:
In Latin America and the Caribbean (LAC), hypertension is the leading cause of death and disability. Accurate diagnosis requires validated blood pressure measurement devices (BPMD). This study aimed to assess the characteristics and validation status of blood pressure measuring devices available for online purchase within Peru.
Methods:
Between September and November 2023, we performed an electronic search of major retailers and online pharmacies. Devices were classified as validated if they were listed in databases recommended by the 2023 European Society of Hypertension (ESH) Guidelines for the Management of Arterial Hypertension or deemed equivalent by the manufacturer. Validation status was verified using a proposed algorithm.
Results:
We identified 13 retailers and 12 pharmacies selling BPMDs online, yielding 530 devices. Of these, 176 were assessed for validation, with 15 devices (8.5%) considered as validated and two (1.1%) as equivalent. Among validated devices, seven used the ESH 2010 protocol, and four each used the ISO 81060-2:2018 standard [adopted by the Association for the Advancement of Medical Instrumentation (AAMI) and ESH] and the ESH 2002 protocol. Six validated devices were from Omron, with two each from Andon, Beurer and Microlife. Prices ranged from 30.1 USD to 351.5 USD.
Conclusion:
The high prevalence of nonvalidated BPMDs available online in Peru underscores the risk of inaccurate blood pressure readings, potentially compromising patient monitoring and treatment. These findings highlight the need for BPMD education for physicians and patients, and advocate for regulatory policies mandating BPMD validation to improve hypertension management in Peru and potentially the LAC region.
More Related Videos
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Errors occurring during blood pressure monitoring
Several factors...
Pre-Procedural Guidelines for Assessing Blood Pressure
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)