Background:
Low- and middle-income countries (LMICs) are disproportionately affected by trauma, resulting in >5 million deaths annually. An essential treatment for musculoskeletal trauma is external fixation. However, in LMICs, current external fixator assemblies are unaffordable, costing patients upward of $5,000 (USD), leaving LMICs to rely on donations that fail to meet the needs of the patient population.
Methods:
New, affordable external fixator (AEFIX) clamps (SONA Global), designed for use in LMIC settings, were compared with Hoffmann 3 (Stryker Medical) clamps to examine their mechanical efficacy compared with commercially available, industry-standard clamps in the U.S. market. In axial and torsional loading scenarios, mechanical testing was performed at the component level and construct level (uniplanar external fixation of a tibial diaphyseal fracture).
Results:
The AEFIX constructs showed no differences in axial or torsional stiffness compared with Stryker constructs. AEFIX constructs yielded at 434.5 N compared with Stryker constructs at 533.6 N (p = 0.04); however, the pins failed before clamp failure could occur in all AEFIX and Stryker constructs. Under cyclic loading at 100 and 300 N, no differences were observed in construct stiffness or interfragmentary gap change. At 500 N of cyclic loading, the AEFIX and Stryker constructs resulted in median gap changes of -4.3 and -1.1 mm, respectively (p < 0.001), with no difference in % change in construct stiffness (p = 0.281).
Conclusions:
AEFIX clamps were comparable to industry-standard Stryker Hoffman 3 clamps in terms of mechanical properties and effectiveness, suggesting that the AEFIX clamps may provide safe external fracture fixation in the non-weight-bearing patient.
Clinical Relevance:
As a safe and affordable solution, AEFIX clamps provide a foundation for enhancing essential trauma surgery capacity in resource-constrained settings around the globe.
More Related Videos
09:26An Efficient and Reproducible Protocol for Distraction Osteogenesis in a Rat Model Leading to a Functional Regenerated Femur
Published on: October 23, 2017
08:20A Reliable and Reproducible Critical-Sized Segmental Femoral Defect Model in Rats Stabilized with a Custom External Fixator
Published on: March 24, 2019
Related Concept Videos
Mechanical Efficiency of Real Machines
However, in reality, no machine can be truly ideal, and all of them experience some...
Equivalent Couples
Two couples are considered to be equivalent if they produce the same rotational effect on a rigid body. In other words, the two couples have the same magnitude and act in the same direction, causing the same angular displacement or acceleration in the body.
For instance, consider two couples lying in the plane of the page, with one having a pair of equal...
Machines: Problem Solving I
The toggle clamp system is a machine structure consisting of movable, pin-connected multi-force members that form a stabilized system to transmit forces. The...
Machines: Problem Solving II
Design Consideration
The factor of safety is another key...
Fatigue