An Infrared Non-Invasive System for Measuring Blood Glucose: A Primary Study using Serum Samples
Ramin Jokari1, Zahra Mahyari2, Mohammad Javad Moulodi2,3
1Student Research Committee, Shiraz University of Medical Sciences, Shiraz, Iran.
Journal of Biomedical Physics & Engineering
|September 4, 2025
Summary
This study developed a non-invasive infrared glucometer. The device achieved 83% accuracy in Zone A of Clarke
Area of Science:
- Biomedical Engineering
- Medical Devices
- Analytical Chemistry
Background:
- Diabetes mellitus presents a significant global health challenge, projected to affect millions by 2024.
- Effective diabetes management relies on accurate and frequent blood glucose monitoring.
- Non-invasive glucose monitoring offers a less burdensome alternative to traditional methods.
Purpose of the Study:
- To develop and evaluate a novel non-invasive glucometer utilizing infrared technology.
- To assess the accuracy of the developed device for blood glucose level measurement.
- To provide a potential tool for improved diabetes patient care and management.
Main Methods:
- Development of a glucometer prototype incorporating infrared transmitter/receiver, power supply, and signal processing.
- Calibration of the device using standard sugar solutions of known glucose concentrations.
- Validation of accuracy using diabetic patient serum samples analyzed with Clarke's grid analysis.
Main Results:
- The non-invasive glucometer demonstrated high accuracy, with 83% of serum samples falling within Zone A of Clarke's grid.
- The remaining 17% of samples were classified in Zone B, with no samples in Zones C, D, or E.
- The device showed enhanced accuracy for glucose concentrations exceeding 200 mg/dL.
Conclusions:
- The developed infrared glucometer shows promise for accurate, non-invasive blood glucose monitoring.
- The high percentage of results in Zone A indicates significant potential for clinical application.
- Further research is recommended to validate the device in larger populations and diverse clinical settings.