Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
Non-invasive PCO2 monitoring during polysomnography: transcutaneous versus end-tidal measurement
Hannah Park1, Jin Yoo1, Ji Su Kim2
1Department of Otorhinolaryngology - Head and Neck Surgery, Ajou University School of Medicine, 164 Worldcup Street, Yeongtong-gu, Suwon, 16499, Korea.
Purpose:
The objectives of this study were to monitor transcutaneous and end-tidal partial pressures of CO2 simultaneously during polysomnography, determine the advantages and disadvantages of each method, and identify relevant factors that can affect the results.
Methods:
This cross-sectional study enrolled 55 adults who underwent polysomnography at the Ajou University Hospital Sleep Center between February 2021 and September 2022. They were volunteers who spontaneously breathed room air. Polysomnography reports, including those of CO2 monitoring, of all participants were reviewed and analyzed by sleep experts. Generalized Estimation Evaluation, Bon Ferroni Post Hoc and multivariable regression analysis were used for statistical analysis.
Results:
Throughout all sleep stages, the mean, highest, and lowest values of end-tidal and transcutaneous partial pressure of CO2 showed significant differences. The mean transcutaneous partial pressure was higher than the mean end-tidal partial pressure by 2.53 mmHg. The apnea index, apnea-hypopnea index, and height were significant factors affecting the difference between the mean transcutaneous and end-tidal partial pressure of CO2. As the obstructive sleep apnea grade increased, the mean end-tidal CO2 partial pressure value decreased. Two patients had hypoventilation; one met the criteria based on the transcutaneous partial pressure of CO2 and the other met those based on the end-tidal partial pressure of CO2.
Conclusion:
During diagnostic sleep studies, the application of both transcutaneous and end-tidal measurements is suggested for stable and accurate monitoring of partial pressure of CO2 and complementary analysis.
More Related Videos
07:52Expired CO2 Measurement in Intubated or Spontaneously Breathing Patients from the Emergency Department
Published on: January 29, 2011
07:54Drug-Induced Sleep Endoscopy DISE with Target Controlled Infusion TCI and Bispectral Analysis in Obstructive Sleep Apnea
Published on: December 6, 2016
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Diagnosing Acidosis and Alkalosis
First, the pH level is assessed to determine whether the blood pH is normal (7.35–7.45), low (acidosis), or high (alkalosis).
Next, the PCO2 and...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...