Concordance study between PowerPlex® ESI 16 Fast Kit and Investigator® 24plex GO! Kit
Yaara Levav Cohen1, Avigail Kleiman1, Aya Spitzer2
1DNA Database Laboratory, Division of Identification and Forensic Science (DIFS), Israel Police, Jerusalem, Israel.
This study compared two short tandem repeat (STR) kits for human identification. The Investigator®24plex GO! Kit showed high concordance with the PowerPlex® ESI 16 Fast Kit, confirming its reliability for forensic databases.
Area of Science:
- Forensic Science
- Human Identification
- Genetics
Background:
- Short tandem repeat (STR) analysis is crucial for forensic human identification.
- Comparing STR multiplex kits is essential for data consistency and database transitions.
Purpose of the Study:
- To evaluate allele call concordance between the PowerPlex® ESI 16 Fast and Investigator®24plex GO! Kits.
- To assess the reliability of the Investigator®24plex GO! Kit for forensic applications.
Main Methods:
- A concordance study using 2000 buccal samples on FTA cards.
- Re-analysis of samples typed with PowerPlex® ESI 16 Fast using the Investigator®24plex GO! Kit.
- Discordance investigation via repeat amplification and sequencing.
Main Results:
- An overall concordance rate of 99.982% was achieved.
- Eleven discordant alleles were identified, including allelic dropout, allele shifts, and one mismatch.
- All discordances were confirmed through repeat testing and sequencing.
Conclusions:
- The Investigator®24plex GO! Kit demonstrates high reliability and concordance with the PowerPlex® ESI 16 Fast Kit.
- This supports the use of the Investigator®24plex GO! Kit for forensic reference databasing.
- A confident transition between these STR typing kits is feasible.
More Related Videos
05:53Candidate Gene Testing in Clinical Cohort Studies with Multiplexed Genotyping and Mass Spectrometry
Published on: June 21, 2018
05:58Digital Polymerase Chain Reaction Assay for the Genetic Variation in a Sporadic Familial Adenomatous Polyposis Patient Using the Chip-in-a-tube Format
Published on: August 20, 2018
Related Concept Videos
Conduction, Convection and Radiation: Problem Solving
In order to solve a problem related to heat transfer, first of all, the situation needs to be examined to determine the type of heat transfer involved. This could...
Electrical Conductivity
In a practical conductor, an applied electric field may be sustained, causing a flow of electrons, which produce a current. The differential form of the current, the current density, is related to the electric field.
More generally, it is related to the force per unit charge, which involves the...
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...
Conduction System of the Heart
This system relies on the unique properties of nodal and Purkinje cells:...