Calibrating Dexcom G7 improves its performance in the context of repetitive recreational scuba diving in people with
Elena Gamarra1, Giovanni Careddu2, Andrea Fazi3
1Servizio di Endocrinologia e Diabetologia, Repubblica e Cantone Ticino Ente Ospedaliero Cantonale, Bellinzona, TI, Switzerland.
Background:
The use of continuous glucose monitors (CGM) in scuba diving for patients with type 1 diabetes (T1DM) shows potential but faces challenges related to accuracy. Previous research has highlighted the poor accuracy of the Dexcom G7 (DG7) in repetitive diving contexts. This study investigates the impact of calibration on the accuracy of DG7, providing valuable insights for patients and clinicians.
Materials And Methods:
In August 2024, 'Diabete Sommerso' organised a 4-day diving cruise around Elba Island (Italy) with 15 participants, including individuals with T1DM. Each participant with diabetes wore two DG7 sensors (one on the arm and one on the abdomen), calibrated daily and compared the results to capillary glucose (Beurer GL50Evo as the reference). Accuracy was assessed using mean absolute relative difference (MARD)/median ARD, Food and Drug Administration (FDA) integrated continuous glucose monitoring (iCGM) criteria and Surveillance Error Grid (SEG) analysis. Hypoglycaemia detection and trends were also evaluated.
Results:
Eight participants with T1DM completed the study using 16 DG7 sensors with no detachments or skin reactions. Analysis of 765 sensor-capillary glucose pairs during 68 dives showed an overall MARD of 13.7%, with arm sensors (11% MARD) outperforming abdomen sensors (16%, p=0.0001). SEG analysis revealed that more than 97% of readings fell within the no-risk zone; however, the FDA's iCGM criteria for non-adjunctive use were not met.
Conclusions:
Calibration improved the accuracy of DG7 in repetitive diving for patients with T1DM. However, capillary glucose checks remain essential, as non-adjunctive criteria were not met.
More Related Videos
15:00Setup of Consumer Wearable Devices for Exposure and Health Monitoring in Population Studies
Published on: February 3, 2023
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...