Left Ventricular Motion Analysis Framework for the MATRIX-VT Study
Christian Janorschke1, Sorin S Popescu2,3, Jonas Osburg4
1Institute of Robotics and Cognitive Systems, University of Lübeck, Ratzeburger Allee 160, Lübeck, 23652, Germany. ch.janorschke@uni-luebeck.de.
Purpose:
Ultrasound (US) is commonly used to assess left ventricular motion for examination of heart function. In stereotactic arrhythmia radioablation (STAR) therapy, managing cardiorespiratory motion during radiation delivery requires representation of motion information in computed tomography (CT) coordinates. Similar to conventional US-guided navigation during surgical procedures, 3D US can provide real-time motion data of the radiation target that could be transferred to CT coordinates and then be accounted for by the radiation system. A motion analysis framework is presented that covers all necessary components to capture and analyse US motion data and transfer it to CT coordinates.
Methods:
Utilizing a robotic test set-up with a human phantom, a baseline and ground truth dataset is recorded for the development and implementation of the motion analysis framework. An optical tracking system and an additional spatial calibration phantom are used to determine necessary transformations. Methods for frame matching, calibration, registration and evaluation are implemented.
Results:
The hardware set-up meets all requirements, including a frame rate exceeding 20 Hz and acceptable image quality, while involving only a few components that can easily be mounted and dismantled in a clinical context.The recorded phantom dataset meets all hardware-specific requirements including a frame rate exceeding 20 Hz, an offset between CT trigger time and the closest US recording of 2-20 ms as well as acceptable US image quality. The static phantom allows for quantitative evaluation by matching structures from different US frames in CT coordinates. While each individual step of the US and CT fusion process achieves the target accuracy of less than 5 mm error, the cumulative error over all transformations exceeded this limit for extreme probe positions.
Conclusion:
The framework is developed and tested for the MATRIX-VT study and can be utilized for patient data evaluation as well as for transferring information such as positional data of moving anatomical structures between US and CTpredictive motion management in STAR therapy. Its modular design allows for the incorporation of advanced calibration and registration methods to address probe positioning limitations, thereby enhancing overall system performance for future applications.
More Related Videos
08:10Estimating Bilateral Atrial Function by Cardiovascular Magnetic Resonance Feature Tracking in Patients with Paroxysmal Atrial Fibrillation
Published on: July 20, 2022
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...
Mitral Regurgitation I: Introduction
Mitral Stenosis II: Clinical features and Diagnostic Tests
Acute Coronary Syndrome III: Diagnostic Studies