Correlations between variability of preoperative ocular biometric measurements and accuracy of intraocular lens power calculation
Related Experiment Video
Updated: Sep 8, 2025

Assessing Early Stage Open-Angle Glaucoma in Patients by Isolated-Check Visual Evoked Potential
Published on: May 25, 2020
Zhenqiang Lin1, Ling Jin2, Yue Wu2
1Department of Ophthalmology, The People's Hospital of Baoan Shenzhen, Shenzhen 518000, China; Department of Ophthalmology, The Second Affiliated Hospital of Shenzhen University, Shenzhen 518000, China; Department of Ophthalmology, Shenzhen Bao'an Air Sea Hospital, Shenzhen 518000, China.
Objectives:
To investigate the relationship between variability of preoperative ocular biometric measurements and the accuracy of intraocular lens (IOL) power calculation.
Methods:
This retrospective study included cataract patients who underwent surgery at the People's Hospital of Baoan Shenzhen from January 2022 to December 2024. Preoperative ocular biometric parameters were measured using the Lenstar LS-900, and the coefficients of variation (CV) of corneal curvature (K1-CV, K2-CV), axial length (AL-CV), anterior chamber depth (ACD-CV) were calculated. Intented residual refractive error (IRRE) was recorded. Postoperative spherical equivalent (SE), prediction error (PE = SE - IRRE), and absolute prediction error (APE = |PE|) were evaluated at one month postoperatively. The mean, median and interquartile range of PE and APE were calculated. Patients were divided into two groups based on the number of valid axial length measurements (Group 1: <3; Group 2: ≥3). Differences in age, sex, preoperative visual acuity (LogMAR VA), K1-CV, K2-CV, AL-CV, ACD-CV, PE, and APE were compared between the two groups. Spearman correlation was used to assess the relationship between biometric variability (K1-CV, K2-CV, AL-CV, ACD-CV) and refractive outcomes (PE and APE).
Results:
A total of 118 eyes were included (Group 1: 44 eyes; Group 2: 74 eyes). Preoperative LogMAR VA was significantly worse in Group 1 than those in Group 2 (0.84 vs. 0.46, p < 0.001), but no significant differences were observed in age, sex, ACD, AL, mean keratometry (Km), PE, or APE between the two groups. Group 1 showed a mean PE of -0.13 and mean APE of 0.61, with 27.03 %, 39.19 %, and 50.0 % of eyes within ±0.25 D, ±0.50 D, and ±1.00 D of target refraction, respectively. In comparison, Group 2 demonstrated a lower mean PE (-1.80) and APE (0.48), and higher proportions of eyes within ±1.00 D (89.19 %). Variability in biometric parameters did not differ between the two groups (p ≥ 0.05). A weak correlation was found between preoperative VA and APE in Group 1 (r = 0.314, p = 0.038), but no significant correlations were detected between variability in biometric parameters with PE or APE.
Conclusions:
The number of valid axial length measurements did not significantly affect preoperative ocular biometric variability or IOL power calculation accuracy in patients with cataract. Patients with fewer valid measurements (<3) had worse preoperative VA, which was weakly correlated with APE.