Can malaria rapid diagnostic tests be used to detect simian malaria?
Wanassanan Madmanee1, Nicholas P J Day2, Nicholas J White2
1Department of Molecular Tropical Medicine and Genetics, Faculty of Tropical Medicine, Mahidol University, Bangkok 10400, Thailand.
Human malaria rapid diagnostic tests (RDTs) show variable performance for simian malaria. Sensitivity improves with higher parasite densities, with some RDTs exceeding 90% detection for key Plasmodium species in non-human primates.
Area of Science:
- Tropical Medicine
- Parasitology
- Diagnostic Development
Background:
- Zoonotic malaria from non-human primates (NHP) presents diagnostic challenges.
- Human malaria rapid diagnostic tests (RDTs) have not been evaluated for simian malaria.
- Accurate diagnostics are crucial for understanding and controlling NHP malaria transmission.
Purpose of the Study:
- To evaluate the performance of commercially available human malaria RDTs in detecting Plasmodium species in non-human primates.
- To assess the impact of parasite density on RDT sensitivity for simian malaria.
- To compare the efficacy of RDTs detecting different Plasmodium antigens (PfHRP2, pLDH, aldolase).
Main Methods:
- Analyzed 131 NHP blood samples with confirmed malaria using 14 different RDT brands.
- Included 30 malaria-negative macaque samples as controls.
- Utilized quantitative PCR (qPCR) as the reference standard for parasite detection and quantification.
Main Results:
- RDT sensitivity varied from 34.2% to 94.3% across brands and Plasmodium species (P. cynomolgi, P. inui, P. coatneyi).
- Sensitivity exceeded 90% for best-performing RDTs when parasite densities were >200 parasites/μL.
- PfHRP2-positive results occurred in 1-17% of malaria-infected NHP samples; no false positives in controls.
Conclusions:
- RDT performance for simian malaria is influenced by RDT brand, Plasmodium species, and parasite density.
- Most RDT brands demonstrate good sensitivity for simian Plasmodium detection at parasite densities above 200 parasites/μL.
- Further research is needed to optimize RDTs for zoonotic malaria surveillance.
More Related Videos
06:18Multiplexed Isothermal Amplification Based Diagnostic Platform to Detect Zika, Chikungunya, and Dengue 1
Published on: March 13, 2018
07:40Field Postmortem Rabies Rapid Immunochromatographic Diagnostic Test for Resource-Limited Settings with Further Molecular Applications
Published on: June 29, 2020
Related Concept Videos
Myasthenia Gravis: Diagnostic Tests
The edrophonium test is a diagnostic tool for myasthenia gravis. It involves...
Pulmonary Tuberculosis IV
Several diagnostic approaches are used to detect TB. The conventional method is the Tuberculin Skin Test (TST), also known as the Mantoux test. However, this method has...