Related Experiment Video
Updated: Jan 18, 2026

In vivo Calcium Imaging in Mouse Inferior Olive
Published on: June 10, 2021
A Case in Which the Hole in the Inflation Line Was Not Closed at the Tip of the Tracheal Tube
Abstract:
We report a rare case in which the inflation lumen at the tip of an endotracheal tube (ETT) was open, leading to intraoperative air leakage and cuff deflation. A patient with Down syndrome undergoing planned dental treatment under general anesthesia was induced and nasally intubated with a cuffed ETT that was then inflated with 5 mL of air. Soon thereafter, it was noted that the pilot balloon was deflated and filled with water droplets. The patient was successfully reintubated with a new, replacement ETT. Upon removal, we examined the defective ETT and sent it onward to the manufacturer. Upon further assessment, the manufacturer reported that the inflation lumen was not properly closed during the manufacturing process because of damage that went undetected. Anesthesia providers should assess an ETT for damage prior to use, including ensuring the cuff is functioning properly.
Related Concept Videos
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Trachea
Anatomical Features:
Location: About half of the trachea is situated in the neck, anterior to the esophagus, and extends from the larynx (at the level of...
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...