Clinical comparison of four commercial MRI phantoms on a 0.35T MR-linac using a standard clinical TRUFI sequence at a
Mateb Al Khalifa1,2, Tianjun Ma1, Haya Aljuaid1
1Department of Radiation Oncology, Virginia Commonwealth University, Richmond, Virginia, USA.
Purpose:
Real‑time magnetic resonance-guided radiation therapy (MRgRT) integrates MRI with a linear accelerator (Linac) for gating and adaptive radiotherapy, which requires robust image‑quality assurance over a large field of view (FOV). Specialized phantoms capable of accommodating this extensive FOV are therefore essential. This study compares the performance of four commercial MRI phantoms on a 0.35 T MR‑Linac.
Methods:
Four phantoms: the IBA QUASAR Insight Phantom, the IBA QUASAR MRID3D Geometric Distortion Phantom, the MagPhan Phantom, and the Sun Nuclear Large Field MRI Distortion Phantom, were evaluated on a 0.35 T MR-Linac using the TRUFI clinical sequence. Mean distortions were measured at gantry angles from 0° to 360° in 30° increments. Clinical imaging parameters (signal-to-noise ratio, uniformity, resolution via modulation transfer function, laser alignment, and twist) were also assessed where applicable.
Results:
The QUASAR MRID3D Phantom enabled separate quantification of B0 and gradient-induced distortions but showed a notable vertical offset, which did not affect comparisons with other phantoms. Both the MagPhan and QUASAR Insight phantoms provided mean distortion measurements as well as additional imaging parameters (e.g., signal-to-noise ratio, uniformity, resolution). The Sun Nuclear Phantom exhibited higher overall distortion values.
Conclusion:
All four phantoms successfully measured distortion at multiple gantry angles on a 0.35 T MR-Linac. The QUASAR MRID3D Phantom uniquely differentiated B0 from gradient distortions but required an offset adjustment. The MagPhan and QUASAR Insight phantoms offered comprehensive imaging metrics, while the Sun Nuclear Phantom exhibited relatively higher distortions.
Related Concept Videos
Imaging Studies IV: Magnetic Resonance Imaging
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies III: Computed Tomography
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...