Related Experiment Video
Updated: Jan 18, 2026

Comprehensive & Cost Effective Laboratory Monitoring of HIV/AIDS: an African Role Model
Published on: October 31, 2010
Point-of-care creatinine-based eGFR (StatSensor) in detecting kidney dysfunction (KD) among people living with HIV in
Daniel Msilanga1,2, Anthony Muiru3, Elizabeth Msangi1,2
1Muhimbili University of Health and Allied Sciences, Dar es Salaam, Tanzania.
Introduction:
Kidney Dysfunction (KD) is prevalent among people living with HIV (PLHIV) in low- and middle-income countries (LMICs), but routine screening is limited due to inadequate laboratory infrastructure. The StatSensor® Point-of-Care (POC) Creatinine Test offers a rapid, cost-effective alternative for early KD detection, though its accuracy in PLHIV remains uncertain.
Methods:
We conducted a diagnostic accuracy cross-sectional study at Temeke Regional Referral Hospital (TRRH) HIV Clinic from January to March 2025 among PLHIV aged ≥18 years. Kidney dysfunction (KD) was defined as an estimated glomerular filtration rate (eGFR) <60 mL/min/1.73 m² using the CKD-EPI 2021 equation. We compared StatSensor point-of-care eGFR results with eGFR derived from serum creatinine measured by the Jaffe method. Diagnostic performance metrics including sensitivity, specificity, predictive values, and receiver operating characteristic (ROC) curves were reported.
Results:
Among 358 participants, the median age was 48 years, with 66.2% female and 15.6% having KD (eGFR < 60 mL/min/1.73m²). The StatSensor demonstrated 92.9% sensitivity, 94.7% specificity, and 94.4% overall diagnostic accuracy compared to the Jaffe method. The ROC curve (AUC = 0.938) indicated strong test performance, showing substantial agreement with a kappa value of 0.805. Bland-Altman analysis revealed a negative bias of 4.36 mL/min/1.73 m² with limits of agreement from -19.68 to 28.40 and a strong correlation (R² = 0.813) between the two methods.
Conclusion:
The StatSensor POC Creatinine test demonstrated high diagnostic accuracy and strong agreement with the standard Jaffe method, indicating its potential as a reliable screening tool for kidney dysfunction in PLHIV in resource-limited settings.
Related Concept Videos
Drug Dosing in Renal Diseases: Estimation of Glomerular Filtration Rate Based on Serum Creatinine Concentration
Drug Dosing in Renal Diseases: Measurement of Serum Creatinine Concentration and Clearance
Factors Affecting Renal Clearance: Renal Impairment
One condition associated with renal failure is uremia. Uremia is characterized by impaired glomerular filtration and fluid accumulation in the body. This condition hinders the renal clearance of drugs, resulting in drug accumulation and potential...
Acute Kidney Injury IV: Diagnostic Studies and Prevention
Pharmacokinetics in Geriatric Patients: Effect of Age on Drug Excretion
Determination of Renal Drug Clearance: Graphical and Midpoint Methods
The graphical method involves plotting the rate of drug excretion in urine against the plasma drug concentration. By analyzing the graph, the clearance can be calculated and obtained. Drugs rapidly excreted by the kidneys exhibit a...