Related Experiment Video
Updated: Jan 18, 2026

06:18
Author Spotlight: Segmentation and VR for Advanced Neurovascular Interventions
Published on: April 5, 2024
1.5K
Workflow Efficiency in Vaginal Cuff High Dose Rate Brachytherapy Using Artificial Intelligence-Based Organ
Yohan A Walter1,2, Lane Rosen1, Olivia Moncrief3
1Department of Radiation Oncology, Willis Knighton Cancer Center, Shreveport, LA 71103, USA.
Cancers
|September 13, 2025
Summary
Artificial intelligence (AI) for organ contouring and applicator modeling significantly reduced high dose rate brachytherapy planning time by 13.7 minutes. These AI tools also decreased inter-user variability in treatment planning, improving workflow efficiency and standardization.
Area of Science:
- Medical physics
- Radiation oncology
- Artificial intelligence in healthcare
Background:
- High dose rate brachytherapy (HDR-BT) is a resource-intensive cancer treatment.
- Workflow inefficiencies and inter-user variability are significant challenges in HDR-BT.
- Artificial intelligence (AI) tools offer potential solutions to these challenges.
Purpose of the Study:
- To evaluate the impact of AI-based organ contouring (AC) and applicator modeling (AM) on workflow efficiency and inter-user variability in vaginal cuff HDR-BT.
- To assess the effect of AC and AM on treatment plan quality.
Main Methods:
- A retrospective analysis of 260 HDR-BT fractions was conducted, comparing cases before and after the implementation of AC and AM.
- Treatment planning times (contouring, dose planning, total) were recorded using timestamps.
- Dosimetric data, including organ-at-risk doses (rectum, bladder D2cc) and target coverage (V90-200), were analyzed.
- Statistical analysis using ANOVA and Tukey tests compared differences between medical physicists before and after AI implementation.
Main Results:
- Implementation of AC and AM led to a significant reduction in overall planning time by 13.7 ± 1.7 minutes.
- Inter-user variability in organ and target dose metrics was reduced across most structures.
- Statistically significant differences in rectum D2cc and target V95 between planners were eliminated after AC and AM implementation.
Conclusions:
- AI-based organ contouring and applicator modeling substantially enhance workflow efficiency in HDR-BT.
- These AI tools maintain or improve treatment plan quality while reducing inter-user variability.
- Standardization of workflows through AI can aid in user training and robust outcome assessment for brachytherapy.
Keywords:
artificial intelligencebrachytherapyhigh dose ratequality improvementradiation oncologyradiation therapytreatment planningvaginal cuff brachytherapyworkflow efficiency