Comparative analysis of pupillometry in two scheimpflug-based measurement systems
Kursat Atalay1, Ibrahim Kocak2, Nihat Sayin2
1Kanuni Sultan Süleyman Training and Research Hospital, Eye Clinic, Atakent Mah. Turgut Ozal Blv. Kucukcekmece, 34303, Istanbul, Turkey. drkursatalay@gmail.com.
Documenta Ophthalmologica. Advances in Ophthalmology
|September 13, 2025
Summary
The Sirius™ and AL-Scan™ devices show comparable mesopic pupillometry (MP) measurements, but photopic pupillometry (PP) results are inconsistent. Further research is needed for device selection in repeated pupillometry.
Area of Science:
- Ophthalmology
- Optometry
- Medical Device Technology
Background:
- Accurate pupillometry and keratometry are crucial for ophthalmic assessments and refractive surgery planning.
- Comparing measurement devices under various conditions ensures reliable clinical data.
- The Sirius™ and AL-Scan™ are modern devices used for ocular measurements.
Purpose of the Study:
- To compare pupillometry measurements (photopic and mesopic) between the Sirius™ and AL-Scan™ devices.
- To evaluate and compare the topographic (keratometric) measurements from both devices.
- To assess device agreement under different lighting conditions.
Main Methods:
- A prospective study included 45 non-smoking patients aged 18-45 years with spherical equivalents between +3.00 D and -5.00 D.
- Pupillometry was measured after 1 minute of dark adaptation.
- Keratometry (K1, K2, KAvg) and pupillometry (photopic and mesopic) were compared between the Sirius™ and AL-Scan™ devices.
Main Results:
- Mesopic pupillometry (MP) showed good correlation between Sirius™ and AL-Scan™ (p=0.65).
- Keratometric values (K1, K2, KAvg) and photopic pupillometry (PP) showed no significant agreement between the devices (p<0.05).
- AL-Scan™ average K1, K2, Kavg: 42.99±1.50 D, 44.04±1.74 D, 43.58±1.62 D.
- Sirius™ average K1, K2, Kavg: 42.94±1.51 D, 43.90±1.75 D, 43.41±1.61 D.
- Average PP: AL-Scan™ 3.97±1.08 mm, Sirius™ 5.05±1.04 mm.
- Average MP: AL-Scan™ 6.04±1.08 mm, Sirius™ 6.01±1.08 mm.
Conclusions:
- The Sirius™ and Nidek AL-Scan™ devices provide comparable mesopic pupillometry data.
- Photopic pupillometry measurements are inconsistent between the two devices.
- Further investigation is required for optimal device selection when repeated pupillometry data is needed.