Related Experiment Video
Updated: Jan 17, 2026

Radiosynthesis, Quality Control, and Small Animal Positron Emission Tomography Imaging of 68Ga-Labelled Nano Molecules
Published on: October 4, 2024
Comparative evaluation of Ordered Subset Expectation Maximization and Bayesian Penalized Likelihood algorithms for
Arvin Naeimi1,2, Sara Harsini3, Ramin Akbarian Aghdam4
1Department of Radiology, McGill University Health Center, McGill University, Montreal, QC, Canada.
Purpose:
This study compares the Ordered Subset Expectation Maximization (OSEM) and Bayesian Penalized Likelihood (BPL) algorithms for Positron Emission Tomography/Computed Tomography (PET/CT) image reconstruction using 18F- fluorodeoxyglucose (FDG) and 68Ga-labeled Prostate-Specific Membrane Antigen(68Ga-PSMA-11) tracers.
Methods:
A retrospective analysis was conducted on 33 patients with various malignancies, including 25 undergoing 18F-FDG PET/CT scans and 8 undergoing 68Ga-PSMA-11 PET/CT scans. Scans were reconstructed using both OSEM and BPL (Q.Clear) algorithms, evaluating key metrics such as Standardized Uptake Value (SUV)max, SUVpeak, background SUV, and tumor-to-background ratio (TBR).
Results:
Thirty-three patients (mean age: 67.53 ± 11.78 years) with 100 lesions (80 FDG, 20 PSMA) were analyzed. In the FDG group, significant differences were observed in lesion SUVpeak, liver SUVpeak, SD of liver SUVmean, bladder SUVmean, SD of bladder SUVmean, and TBR, with BPL generally producing higher values except for liver SUVpeak, SD of liver SUVmean, and SD of bladder SUVmean. In the PSMA group, BPL enhanced most metrics except for the SD of liver SUVmean and the SD of bladder SUVmean. While strong linear correlations between BPL and OSEM metrics were observed (Pearson r>0.85 for most parameters), Bland-Altman analysis revealed wide limits of agreement, particularly for TBR in the PSMA group (-13.00 to 23.89), indicating substantial variability in absolute values between methods. Assessment of the relationship between lesion volume and SUVmax differences (BPL-OSEM) revealed a weak, non-significant negative correlation in the total cohort(r = - 0.14, p = 0.16) and in the FDG subgroup(r= - 0.14, p = 0.24).
Conclusion:
Both reconstruction methods demonstrate clinical utility, with BPL producing statistically higher values for several quantitative metrics, such as SUVmax and TBR, without markedly improving lesion detectability. While strong correlations were observed between BPL and OSEM values, the wide limits of agreement, particularly for TBR in PSMA imaging, suggest these methods may not be directly interchangeable in longitudinal studies. Harmonization strategies may help reduce inter-method variability and improve scan comparability in longitudinal or multicenter settings. Prospective approaches, such as reconstruction-specific reference ranges or scaling factors, could further support harmonization efforts in clinical trials. For longitudinal monitoring, consistent use of the same reconstruction method is recommended to ensure reliable quantification.
More Related Videos
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
10:28Gene Regulation and Targeted Therapy in Gastric Cancer Peritoneal Metastasis: Radiological Findings from Dual Energy CT and PET/CT
Published on: January 22, 2018
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET