Real-world effectiveness of DKutting Scoring Balloon for AVF and AVG stenosis and thrombosis
Lihong Zhang1, Fan Zhang1, Xiangru Li1
1Department of Nephrology, Haidian Hospital, Beijing, China.
Insights
The DKutting Scoring Balloon shows promising results for treating arteriovenous fistula and graft stenosis in hemodialysis patients, achieving high patency rates and demonstrating safety. Further trials are recommended to confirm its effectiveness.
Area of Science:
- Vascular Surgery
- Interventional Radiology
- Nephrology
Background:
- Arteriovenous fistula (AVF) and arteriovenous graft (AVG) stenosis are frequent complications in hemodialysis patients, often leading to restenosis and requiring repeat interventions.
- Conventional percutaneous transluminal angioplasty (PTA) has limitations in achieving long-term efficacy for these stenotic lesions.
Purpose of the Study:
- To evaluate the safety and efficacy of the DKutting Scoring Balloon in treating AVF and AVG stenosis.
- To assess the primary patency rates and adverse events associated with the DKutting Scoring Balloon.
Main Methods:
- Retrospective analysis of 428 patients with AVF or AVG stenosis treated with the DKutting Scoring Balloon at Beijing Haidian Hospital.
- Evaluation of primary patency rates at 1, 3, and 6 months post-intervention.
- Subgroup analysis for AVF and AVG, including resistant and recurrent cases.
Main Results:
- High primary patency rates were observed: 98.1% at 1 month, 90.8% at 3 months, and 78.0% at 6 months.
- Six-month patency rates were 79.9% for AVF and 75.4% for AVG.
- No major adverse events were reported, indicating a favorable safety profile.
Conclusions:
- The DKutting Scoring Balloon demonstrates potential for improved outcomes in treating AVF and AVG stenosis compared to conventional PTA.
- The device appears safe and effective, even in challenging cases of resistant and recurrent stenosis.
- Further randomized controlled trials are warranted to validate these findings and guide clinical practice.
Abstract:
Arteriovenous fistula (AVF) and arteriovenous graft (AVG) stenosis are common complications among hemodialysis patients, necessitating repeated interventions due to neointimal hyperplasia and restenosis. Percutaneous transluminal angioplasty (PTA), while a standard treatment, faces challenges in long-term efficacy. The DKutting Scoring Balloon, a novel advancement in scoring balloon technology, incorporates triangular scoring elements and non-compliant balloon material to create controlled micro-incisions in the vessel wall, aiming to enhance procedural precision, reduce vessel trauma, and lower restenosis rates. In the largest study of its kind, this retrospective analysis conducted at Beijing Haidian Hospital evaluated the device's performance across 428 patients with significant AVF and AVG stenosis. The findings indicate primary patency rates of 98.1%, 90.8%, and 78.0% at 1, 3, and 6 months, respectively. Subgroup analyses revealed patency rates of 79.9% for AVF and 75.4% for AVG at 6 months, demonstrating its effectiveness even in resistant and recurrent cases. Importantly, no major adverse events were reported, underscoring the device's safety profile. These observational findings suggest that the DKutting Scoring Balloon may offer favorable outcomes compared to those typically reported with conventional PTA, though these results should be interpreted as hypothesis-generating and further randomized controlled trials are recommended to corroborate these findings and optimize its integration into clinical practice.