Related Experiment Video
Updated: Jan 17, 2026

Hollow Microneedle-based Sensor for Multiplexed Transdermal Electrochemical Sensing
Published on: June 1, 2012
Microneedle-assisted interstitial fluid extraction combined with a 2D tellurene-based glucose sensor for personal
Isha Basumatary1, Shreya Shashank Chauhan2, Venkata Vamsi Krishna Venuganti2
1Department of Electrical and Electronics Engineering, BITS Pilani, Hyderabad Campus, Hyderabad, 500078, India. parikshit@hyderabad.bits-pilani.ac.in.
None:
The objective of this work is to develop a selective glucose sensor for real-time in vivo monitoring of interstitial fluid (ISF) glucose levels, addressing challenges of interference from other biomolecules. A microneedle (MN) patch was employed for ISF extraction, while a 2D tellurene (Te)-based glucose oxidase (GOX) GOX/Te/carbon cloth (CC) electrode served as the sensing element for electrochemical measurements. The sensor exhibited a low detection limit (LOD) of 0.357 μM, and a sensitivity of 0.10378 mA μM-1 cm-2. In vivo studies in an animal model demonstrated an average glucose concentration of 5.05 mM with a rapid response time of ∼95 s. Importantly, the electrochemical measurements showed a strong correlation with commercial glucometer readings, confirming accuracy and reliability. This study highlights the potential of the proposed sensor for real-time in vivo glucose monitoring, paving the way for future home care diabetes management devices.