Near-Zero Fluoroscopy Ablation Workflow With a Circular Multielectrode Pulsed-Field Ablation Catheter
Kennosuke Yamashita1, Yohei Kikuchi1, Keita Yoshiyama1
1Department of Cardiovascular Medicine, Heart Rhythm Center, Sendai Kosei Hospital, Miyagi, Japan.
Pacing and Clinical Electrophysiology : PACE
|September 16, 2025
Summary
Pulsed field ablation (PFA) with the PulseSelect catheter significantly reduces radiation exposure during atrial fibrillation (AF) treatment. This near-zero fluoroscopy approach enhances safety and maintains ablation efficacy.
Area of Science:
- Cardiology
- Medical Devices
- Electrophysiology
Background:
- Pulmonary vein isolation (PVI) using pulsed field ablation (PFA) is effective for atrial fibrillation (AF).
- Traditional PVI often requires fluoroscopy, increasing radiation exposure risks.
- A novel workflow using the PulseSelect catheter with 3D electroanatomical mapping (3D-EAM) and intracardiac echocardiography (ICE) aims to minimize fluoroscopy.
Purpose of the Study:
- To evaluate the feasibility and efficacy of a near-zero fluoroscopy PVI workflow for AF treatment.
- To assess the safety and procedural outcomes of using the PulseSelect catheter with 3D-EAM and ICE.
Main Methods:
- Procedures were performed under general anesthesia with transseptal puncture.
- Pre-mapping with a multi-electrode catheter guided visualization of the PulseSelect catheter on 3D-EAM for PVI.
- Post-procedure mapping confirmed isolation and assessed for electrical reconnection.
Main Results:
- Median isolation times were 9.0 min for left and 11.0 min for right pulmonary veins.
- Average total procedure time was 48.5 min, with only 0.1 min of fluoroscopy for initial registration.
- Catheter positioning was controlled with minimal deviation, no significant complications, and a low recurrence rate at three months.
Conclusions:
- The near-zero fluoroscopy PFA workflow using PulseSelect, 3D-EAM, and ICE significantly reduces radiation exposure.
- This approach enhances procedural safety and maintains efficacy for AF ablation.
- The study confirms the feasibility of reducing fluoroscopy reliance while achieving accurate and safe ablation outcomes.