Related Experiment Video
Updated: Jan 17, 2026

Treating Low Back Pain in Failed Back Surgery Patients with Multicolumn-lead Spinal Cord Stimulation
Published on: June 26, 2018
Spinal Cord Stimulator Migration and Malfunction in the Setting of Upper Extremity Complex Regional Pain Syndrome: A
Mohammed Miniato1, Krishna Shah1
1Department of Anesthesiology, Baylor College of Medicine, Houston, TX.
Background:
Common long-term complications for neuromodulation include lead migration, loss of efficacy of therapy, and lead malfunction. The following case describes pitfalls of lead migration and preventative steps to decrease risk of this complication.
Case Report:
This is an independent case study following the outcome of one patient diagnosed with complex regional pain syndrome that had failed conservative therapy. She underwent implantation of a spinal cord stimulator (SCS) with > 90% relief. Loss of relief was described on follow-up. On reimaging, the SCS had migrated cephalad and this was likely the cause of the malfunction.
Conclusions:
Lead migration can occur at any time, even years after initial implantation. Recommendations from the literature and gained from this case include heightened vigilance and suspicion when a patient encounters sudden failure of pain relief, low threshold for new imaging to assess lead location, and use of mechanical anchoring devices.