X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Patient Dose Optimisation of Abdominopelvic Protocols During X-Ray Imaging
Emmanuel Ekem-Ferguson1, Shirazu Issahaku1,2, Mark Pokoo-Aikins1,2
1Department of Medical Physics, University of Ghana, Legon, Accra, Ghana.
Introduction:
Abdominopelvic radiography is commonly performed in both supine and erect positions, raising concerns about radiation dose exposure. In Ghana, there is limited data on organ dose evaluation in these examinations, particularly regarding patient positioning for dose optimisation. Additionally, the risk of dose creep in Digital Radiography (DR) due to its wide dynamic range further complicates dose management. This study aimed to estimate organ doses for patients undergoing abdominopelvic radiography in both positions and to develop an optimisation strategy that minimises radiation exposure while maintaining image quality.
Methods:
Organ doses were estimated using CalDose X 5.0 for 112 patients across two facilities in both erect and supine positions. Image quality was assessed using signal-to-noise ratio (SNR) values for five specific organs, with the kidney, the liver, and the prostate selected for dose and image quality optimisation. Additionally, 34 exposures were performed on an anthropomorphic phantom using varying exposure parameters and focus-to-detector distances (FDD), with thermoluminescence dosimeters (TLDs) employed for dose optimisation measurement.
Results:
An average of 39.9% and 52.9% of radiographs had optimal and high SNR values respectively. After the optimisation procedure, the recommended exposure parameters were 70 kVp, 10 mAs, and 110 cm FDD for both positions. In the supine position, organ doses reduced with a percentage difference of 50.0, 65.22 and 19.21, and 23.03, 28.97 and 30.98 for SNR for the kidney, the liver and the prostate respectively. In the erect position, organ doses reduced with a percentage difference of 72.34, 72 and 4.55, and SNR percentage difference of 10.69, 9.39 and 32.83 for the kidney, the liver and the prostate respectively.
Conclusion:
Abdominopelvic organ doses vary based on patient demographics and exposure settings. The optimisation strategy significantly reduced radiation doses while preserving image quality. These findings emphasise the importance of standardised imaging protocols to enhance patient safety and radiation protection in clinical practice.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
08:34Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
Related Concept Videos
Radiological Investigation I: X-ray and CT
X-ray Imaging
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies III: Computed Tomography