Related Experiment Video
Updated: Jul 10, 2026

Optical Detection of E. coli Bacteria by Mesoporous Silicon Biosensors
Published on: November 20, 2013
Highly Sensitive Dissolved Oxygen Sensor with High Stability in Seawater Based on Silica-Encapsulated Platinum(II)
Hang Lv1,2, Siyuan Cheng1, Song Hu1
1State Key Laboratory of High Performance Ceramics and Superfine Microstructure, Shanghai Institute of Ceramics, Chinese Academy of Sciences, Shanghai 200050, China.
None:
This study utilized tetramethylammonium hydroxide (TMAH) as a substitute for traditional catalysts and successfully incorporated platinum octaethylporphyrin (PtOEP) into SiO2 nanoparticles (PtOEP@SiO2) via the Stöber method. Methyl silicone resin was employed as the matrix material, and a drop-coating technique was applied to fabricate thin films of PtOEP@SiO2 particles for dissolved oxygen (DO) sensing in seawater. By optimizing the concentrations of TMAH and PtOEP, a highly sensitive oxygen-sensing film with a quenching ratio (I0/I100) of 28 was ultimately developed, with a wide linear detection range (0~20 mg/L, R2 = 0.994). Stability tests revealed no significant performance degradation during five oxygen-nitrogen cycle tests. After 30 days of immersion in East China Sea seawater, the quenching ratio decreased by only 6%, confirming its long-term stability and excellent resistance to ion interference. This research provides a novel strategy for developing highly reliable in situ marine DO sensors.