Comparison of steam and hydrogen peroxide gas plasma sterilization of cohesive latex bandage wraps: A pilot study
Sarah L Austin1, Alec Truswell2, Kadie O'Byrne3
1Brindabella Veterinary Referral Services, Holt, Canberra, Australian Capital Territory, Australia.
Objective:
To compare the sterilization of cohesive bandage material with steam or hydrogen peroxide gas plasma (HPGP).
Study Design:
In vitro experimental pilot study.
Sample Population:
Latex cohesive bandages (n = 49).
Methods:
Rip-Rap cohesive bandages (n = 20) in equal groups; as supplied, rewound with or without the cardboard roll, and as supplied with no cardboard roll. These were inoculated with 0.1 mL Geobacillus stearothermophilus spores. Each was packaged with a chemical and biological indicator before sterilization with HPGP. Wrap.it (n = 10) and Rip-Rap (n = 10) bandages, divided into as-supplied and rewound groups, were inoculated with 0.1 mL G. stearothermophilus spores. Then sterilized in a steam autoclave at 134°C for 5 min. Post sterilization, the inoculated area was resected and placed into broth and incubated at 60°C for 48 h (HPGP) and 24 h (Steam), then plated onto nutrient agar plates and incubated at 60°C for 24 h. A positive result was recorded as any growth and confirmed as a pure G. stearothermophilus population via MALDI-TOF.
Results:
A total of 16 of the HPGP samples had growth of G. stearothermophilus post sterilization. A total of 12 chemical indicators and four biological indicators passed. None of the steam sterilized samples had positive growth, and all indicators passed.
Conclusion:
HPGP is not an adequate method of sterilization for latex cohesive bandages. Steam sterilization is effective when the critical parameters for length and temperature are met.
Clinical Significance:
Evidence-based recommendations guide in-house sterilization of latex cohesive bandages for use within the surgical field. This is an important aspect of draping aseptically and, therefore, infection prevention.
Related Concept Videos
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Hand hygiene
Hand washing...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Chemical Agents for Microbial Control
Physical Methods for Controlling Microbial Growth: Temperature