Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
Transcutaneous versus arterial carbon dioxide monitoring in adult polysomnography studies
Keeley A Miller1, Thomas J Churchward2,3, Julie Tolson1,2,3
1Department of Medicine, University of Melbourne, Melbourne, VIC, Australia.
Study Objectives:
To determine the accuracy of transcutaneous carbon dioxide measurements using SenTec tcPCO2 during adult polysomnography (PSG) compared to arterial PCO2.
Methods:
In consecutive patients having PSG with transcutaneous CO2 (tcPCO2) monitoring using a SenTec monitor, arterial blood gas samples were taken at the beginning and at the end of each sleep study. tcPCO2 measurements recorded at 0, 60, and 120 s after arterial sampling was determined in 51 participants who underwent PSG with tcPCO2 monitoring at the Austin Health Sleep Laboratory (Heidelberg, VIC, Australia) during the period 01/05/22-10/06/22. The mean of differences between arterial carbon dioxide tension (PaCO2) and the predicted value based on tcPCO2 measurements determined at each time point for evening and morning arterial samples.
Results:
Data were obtained from 37 participants. A statistically significant difference in the tcPCO2 values at the point of arterial sampling was found, with tcPCO2 on average 2.2-2.8 mmHg less than PaCO2 at various timepoints after arterial sampling. The upper limit of measurement accuracy represented a moderate difference (2-4.9 mmHg), within the clinically acceptable range of -7.5 to +7.5 mmHg variation from PaCO2 values.
Conclusions:
The SenTec tcPCO2 electrode was found to be an appropriate alternative to arterial sampling for assessment of arterial carbon dioxide in adult patients in the PSG setting, with no significant difference in tcPCO2 accuracy based on the timing of transcutaneous measurements (up to 2 min) following arterial sampling.
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Carbon Dioxide Transport in the Blood
Forms of CO2 Transport
1. Dissolved in plasma: A small percentage (7-10%) of CO2 is transported and dissolved directly in the plasma.
2. Carbaminohemoglobin: Just over 20% of CO2 is chemically bound to...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...