Reproducibility of diastolic function parameters in repeated ECG-gated SPECT myocardial perfusion imaging and
Aaro Krohns1, Tomi P Laitinen2,3, Tiina M Laitinen2
1University of Eastern Finland, Kuopio, Finland.
Objective:
To study the reproducibility of diastolic function parameters from myocardial perfusion imaging (MPI) using ECG-gated single photon emission computed tomography (SPECT), and echocardiography in repeated imaging.
Methods:
We studied the reproducibility of MPI diastolic function parameters peak-filling rate (PFR) and time to peak filling (TTPF) as well as echocardiographic parameters E/A, E/e', and E-wave deceleration time (DT). The study population consisted of 21 patients who underwent routine MPI with repeated rest acquisition and echocardiography. In a test-retest setting, appropriate diastolic measures were successfully obtained using SPECT in 20 patients, with E/A measured in 18, E/e' in 17, and DT in 16 patients.
Results:
We found that PFR was well reproducible with the intraclass correlation coefficient (ICC) 0.887, and coefficient of variation (CV%) 10.5. However, TTPF was poorly reproducible (ICC 0.012, and CV% 17.5). E/A was highly reproducible (ICC 0.963, and CV% 12.5). Decent reproducibility was observed for E/e' (ICC 0.809, and CV% 18.6) and DT (ICC 0.833, and CV% 12.0). We further studied the correlation between these parameters. PFR (EDV/s) had negative correlation with DT (r = -0.538, p = 0.039) and E/A had positive with E/é (r = 0.689, p = 0.002). No other significant correlations were observed (p > 0.05 for all). We also examined how parameters classified patients as normal or abnormal regarding the diastolic function. E/A and E/e´, as well as E/A and PFR showed similar classifications in 88% of cases, with a Kappa value of 0.433, p = 0.074 for both.
Conclusions:
PFR obtained from repeated SPECT studies, along with the E/A ratio, is highly reproducible.
More Related Videos
11:04Quantification of Global Diastolic Function by Kinematic Modeling-based Analysis of Transmitral Flow via the Parametrized Diastolic Filling Formalism
Published on: September 1, 2014
12:12Echocardiographic Approaches and Protocols for Comprehensive Phenotypic Characterization of Valvular Heart Disease in Mice
Published on: February 14, 2017
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...