Focus correction in MR thermography for increased targeting precision during focused ultrasound procedures
Chang-Sheng Mei1,2, Shenyan Zong3,4, Bruno Madore2
1Department of Physics, Soochow University, Taipei, Taiwan.
Purpose:
Accurate targeting during MR-guided focused ultrasound (FUS) procedures is essential for effective treatment to be achieved. However, spatial discrepancies frequently arise between the planned target and the observed thermal hotspot on proton resonance frequency (PRF)-based MR thermometry because of temperature-induced artifacts. This study aims to correct such displacements caused by chemical shift and k-space center offset.
Methods:
Spatial misregistration was addressed using a two-step correction approach. The first step corrected pixel-wise displacements attributed to temperature-dependent resonance frequency shifts (chemical shift), based on local frequency offset maps. The second step compensated for TE errors induced by asymmetric phase gradients near the thermal focus, restoring accuracy in hotspot localization. Validation was performed in controlled phantom experiments, and the approach was retrospectively tested in vivo, in 121 sonications across seven essential tremor (ET) patients.
Results:
Phantom experiments demonstrated that spatial shifts up to approximately 1.5 mm could be effectively corrected. Clinical analysis showed a strong correlation (R2 = 0.852) between temperature rise and spatial displacement, with a mean shift of 0.5 Mm per 10°C. Combined correction significantly reduced temperature estimation bias, with the mean error decreasing from -0.11°C to -0.05°C, as evaluated by Bland-Altman analysis.
Conclusion:
Temperature-related chemical shift and k-space offset substantially impact the spatial fidelity of PRF-based MR thermometry. The proposed correction framework improves thermal hotspot localization, enabling more accurate lesion targeting during FUS procedures.
More Related Videos
08:08Evaluating Targeting Accuracy in the Focal Plane for an Ultrasound-guided High-intensity Focused Ultrasound Phased-array System
Published on: March 6, 2019
13:41Magnetic Resonance-Guided High Intensity Focused Ultrasound Generated Hyperthermia: A Feasible Treatment Method in a Murine Rhabdomyosarcoma Model
Published on: January 13, 2023
Related Concept Videos
Atomic Nuclei: Magnetic Resonance
NMR Spectrometers: Resolution and Error Correction
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Super-resolution Fluorescence Microscopy
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...