Diffuse Reflectance Spectroscopy: Getting the Capillary Refill Test Under One's Thumb
Published on: December 2, 2017
Examining intra- and inter-device reliability of pressure-mediated reflection spectroscopy in a multi-state sample of
Susan B Sisson1, Emily Helms1, Shanon Casperson2
1Department of Nutrition Sciences, University of Oklahoma Health Sciences Centerhttps://ror.org/0457zbj98 , Oklahoma City, OK, USA.
Objective:
To examine the intra- and inter-device reliability of devices using pressure-mediated reflection spectroscopy (the Veggie Meter®).
Design:
A cross-sectional research study was conducted across eight sites in the USA. Using two Veggie Meters® at each site, participants completed five, counter-balanced pairs of finger scans. Intra-device comparisons included intra-class correlation coefficients (ICC) and calculation of the CV and 95 % CI of each device/site; hypothesised to be ≤ 6 %. Inter-device comparisons included ICC, absolute relative differences (ARD) and 95 % CI, and equivalence; both hypothesised to be ≤ 10 %.
Setting:
Eight sites across the USA.
Participants:
Across sites, participants' (n 282) average age ranged 24·7-39·0 years; sex ranged 60·0-85·7 % women and Non-Hispanic White ranged 20·0-94·3 %.
Results:
Intra-device ICC ranged from 0·77 to 0·99. The CV ranged from 6·2 to 14·2 %, with an average of 8·8 %. A majority (63 %; n 10) of the Veggie Meter® devices had significantly higher CV from the hypothesised 6 %. Inter-device ICC ranged from 0·58 to 0·94. The ARD ranged from 7·5 to 22·0 %, with an average of 13·9 %. ARD in a majority (n 5) of sites was significantly higher than the hypothesised 10 %. Five sites (63 %) demonstrated equivalence below the hypothesised 10 %.
Conclusions:
Our study demonstrates the intra-device and inter-device reliability to be moderate to high, as per ICC. The observed margin of difference within a device was up to 14 %, with an average of 9 %. The observed margin of difference between devices was up to 22 %, with an average of 14 % between devices.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...