Related Experiment Video
Updated: Jan 16, 2026

A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Implantable Cardioverter-Defibrillator Therapies Following Generator Replacements-Long-Term Remote Monitoring Data
Maciej Dyrbuś1, Łukasz Pyka1, Anna Kurek1
13rd Department of Cardiology, School of Medical Sciences in Zabrze, Medical University of Silesia, 40-055 Katowice, Poland.
Long-term outcomes after implantable cardioverter-defibrillator (ICD) or cardiac resynchronization therapy-defibrillator (CRT-D) replacement are crucial. This study found no significant link between improved LVEF and reduced device therapies or better survival post-replacement.
Area of Science:
- Cardiology
- Electrophysiology
- Medical Device Technology
Background:
- Long-term outcomes following implantable cardioverter-defibrillator (ICD) and cardiac resynchronization therapy-defibrillator (CRT-D) replacements are not well-established.
- Understanding arrhythmic episodes and mortality risk in remotely monitored patients after device replacement is essential for clinical management.
Purpose of the Study:
- To evaluate the risk of ICD or CRT-D therapies in remotely monitored patients undergoing device replacement.
- To identify predictors of all-cause mortality and appropriate/inappropriate device therapies post-replacement.
Main Methods:
- Analysis of data from 134 patients who underwent ICD/CRT-D replacement or upgrade.
- Utilized Kaplan-Meier estimates and Cox proportional hazards regression for outcome analysis.
- Endpoints included all-cause mortality, and appropriate and inappropriate ICD/CRT-D therapies.
Main Results:
- The cohort comprised 51.5% ICD and 48.5% CRT-D recipients, with a median LVEF of 23.0% at replacement.
- During a median follow-up of 3.0 years, 32.1% experienced appropriate and 6.0% experienced inappropriate therapies.
- All-cause mortality was 38.0%; predictors of death included appropriate antitachycardia pacing (ATP), reduced LVEF, and no history of myocardial infarction.
Conclusions:
- Improvement in left ventricular ejection fraction (LVEF) was not a predictor of improved survival or reduced device therapies after ICD/CRT-D replacement.
- An LVEF >35% at replacement did not influence analyzed outcomes.
- Further stratification models are needed to better evaluate arrhythmic risk in patients post-generator replacement.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
04:24A Novel Digital Platform for a Monitored Home-based Cardiac Rehabilitation Program
Published on: April 19, 2019
Related Concept Videos
Cardiomyopathy V: Interprofessional Care
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Holter Monitor: 24-Hour Monitoring
Cardiomyopathy II: Dilated Cardiomyopathy
Heart Failure VI: Adjunct Therapies
Endocarditis III: Medical Management