Estimation of organ and effective doses for rotational cerebral angiography using the National Cancer Institute
Hitoshi Miyazaki1,2, Toshioh Fujibuchi3, Donghee Han3
1Division of Radiology, Department of Medical Technology, Kyushu University Hospital, 3-1-1 Maidashi, Higashi-ku, Fukuoka, 812-8582, Japan. miyazaki.hitoshi.036@m.kyushu-u.ac.jp.
Abstract:
Rotational cerebral angiography requires accurate dosimetry. The National Cancer Institute Dosimetry System for Radiography and Fluoroscopy (NCIRF), a Monte Carlo-based dosimetry software, can evaluate the organ dose (OD) and effective dose (ED) with higher accuracy than the conventional Monte Carlo software (PCXMC). We estimated the OD and ED for three-dimensional digital subtraction angiography (3D-DSA) and cone beam computed tomography (CBCT) using the NCIRF, reflecting dose variations during rotational cerebral angiography. The 3D-DSA and CBCT simulation parameters were obtained by rotational imaging of a physical head phantom using the Artis Q biplane system. The air kerma area product for each projection was determined based on the ratio of the tube current-time product for each projection; the NCIRF was used with male and female voxel-type reference computational phantoms. To validate the simulation results, the lens dose of the phantom was measured using radiophotoluminescence glass dosimeters and compared to the simulated lens dose. The highest ODs were delivered to the brain: 8.8 mGy (males) and 11.6 mGy (females) in 3D-DSA and 50.0 mGy (males) and 59.4 mGy (females) in CBCT. The EDs were 0.27 mSv (males) and 0.35 mSv (females) in 3D-DSA and 1.49 mSv (males) and 1.83 mSv (females) in CBCT. Lens doses differed within 8.0% between measurements and simulations, with 45.9-65.5% overestimation in simulations that did not account for dose variability. Simulations that considered dose variability using the NCIRF more accurately estimated OD and ED in rotational cerebral angiography.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Biological Effects of Radiation
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...